background image

Air Overlay Mattress User Guide

3

Please read the following instructions carefully and observe the 

warning instructions before using the system.

Contents

03

 

The system

04

 

System set-up

05

 Operation

05

 

System maintenance

06

 Cleaning

06

 

Technical specification

07

 

Troubleshooting guide

08

 

Waste disposal

08

 Warranty

09

 Safety

Edward House, 5 Penner Road, Havant, PO9 1QZ

The system 

OSKA SeriesC Air is a cushion system which is 

designed to be placed on any chair. It is designed 

for use in nursing care at home, or in a nursing 

facility environment.
Application – for use on a chair within home 

or care environment with power source nearby 

to ensure that comfort and relief levels are 

maintained.
The system consists of a control unit featuring a 

membrane pump and a cushion with 6 air cells 

arranged in a transverse manner. The entire cell 

unit is encased in a cradle of visco foam. The 

bielastic top cover and PVC outer base cover, 

ensure stability, comfort and infection control. 

 

Recommendations

OSKA SeriesC Air alternating cushion system is 

suitable as an aid in the prevention and treatment 

of pressure ulcers up to and including category 2. 

 

Capacity

 

OSKA SeriesC Air alternating cushion system is 

recommended for patient weight up to 110kg. 

Heavier patients are better accommodated with a 

more appropriate OSKA pressure system.

Summary of Contents for SeriesC Air

Page 1: ...User Guide SeriesC Air ...

Page 2: ... use on a chair within home or care environment with power source nearby to ensure that comfort and relief levels are maintained The system consists of a control unit featuring a membrane pump and a cushion with 6 air cells arranged in a transverse manner The entire cell unit is encased in a cradle of visco foam The bielastic top cover and PVC outer base cover ensure stability comfort and infectio...

Page 3: ...ns and the products Press the alarm reset button see 4 to mute the audible alarm its indicator will flash 1A Operation 01 When the pressure reaches full inflation within approximately 5 minutes the normal pressure indicator will illuminate and the low pressure indicator will turn off 02 Turn the pressure dial see 5 to set a comfortable pressure level from 40 110kg according to the approximate weig...

Page 4: ...t the appliance to the nearest mains socket and ensure that the power indicator light is on Contact OSKA Service The pump appears to be on running but is not reaching the desired pressure setting The hoses are connected correctly Disconnect the hoses and re connect them in to the outlet connectors on the side of the pump Contact OSKA Service The pump is running but is not inflating the cushion The...

Page 5: ...e user owner WARNING For private non professional operators In the case of system failure please check 01 Is there sufficient power supply 02 Is the unit turned on 03 Is the power cord connected properly and is the power plug inserted correctly 04 Is the mattress connection tube connected properly If all these primary checks have been carried out and the system continues not to function then you m...

Page 6: ...Air Overlay Mattress User Guide Edward House 5 Penner Road Havant PO9 1QZ 0845 381 0600 ask oska uk com oska uk com Pressure Care Experts ...