Air Overlay Mattress User Guide

7

Cleaning

Remove dirt or stains with lukewarm water and a 

cloth. Water may be treated with a mild cleaning 

agent. Do not use aggressive cleaning agents or 

solvents. You must disinfect surfaces regularly, 

at least after every clean. Wipe disinfection is 

preferred. The cleaning agent you use must comply 

with the manufacturers information as stated on 

the cleaning agent. If in doubt you must request 

further information.

Technical specification

 

Pump

OSKA – SeriesC Air

Power supply: 

220~240V/50Hz, 1A

Air output: 

8 liter/min.

Pressure range: 

30 mmHg – 60 mmHg

Cycle time: 

12 min. 

Static Mode 

Alarm Reset Mode 

With normal / low pressure light

Size: 

28cm(L) x 20cm(W) x 10cm(H)

Weight: 

2.2 kg

Cushion

Size: 

49cm(L) x 46cm(W) x 10cm(H)

Top cover:

 Dark blue bi-elastic cover with PU/

Polyester coating and zip fastening

Air cells: 

Nylon/PVC

Base: 

Nylon/PVC with foam insert 6 Air Cells

Troubleshooting guide

The pump appears 

not to be on running

The mains supply is 

plugged in. The switch 

is in the on position. 

The fuse in the plug is 

not blown

Connect the appliance 

to the nearest mains 

socket and ensure that 

the power indicator light 

is on

Contact OSKA Service

The pump appears 

to be on running 

but is not reaching 

the desired pressure 

setting

The hoses are 

connected correctly

Disconnect the hoses 

and re-connect them in 

to the outlet connectors 

on the side of the pump

Contact OSKA Service

The pump is running 

but is not inflating 

the cushion

The hoses are 

connected correctly

Disconnect the hoses 

and re-connect them in 

to the outlet connectors 

on the side of the pump

Contact OSKA Service

There are no leaks in 

the cushion

Contact OSKA Service

Contact OSKA 

Service

There are no kinks in 

the tubing down side 

of the cushion

Carefully untwist any 

kinks

Contact OSKA Service

Check that the 

pressure has been 

set in the correct 

range for the patient’s 

weight

Adjust the pressure 

setting accordingly

Contact OSKA Service

The system does 

not appear to be 

alternating

There are no kinks in 

the tubing down side 

of the cushion

Carefully untwist any 

kinks

Contact OSKA Service

The operation pump 

is noisy

The pump is not 

resting against a solid 

surface

Re-position the pump to 

a suitable surface

Contact OSKA Service

Fault occurring

Ensure that

Possible solution

If your product is still 
not working

Summary of Contents for SeriesC Air

Page 1: ...User Guide SeriesC Air ...

Page 2: ... use on a chair within home or care environment with power source nearby to ensure that comfort and relief levels are maintained The system consists of a control unit featuring a membrane pump and a cushion with 6 air cells arranged in a transverse manner The entire cell unit is encased in a cradle of visco foam The bielastic top cover and PVC outer base cover ensure stability comfort and infectio...

Page 3: ...ns and the products Press the alarm reset button see 4 to mute the audible alarm its indicator will flash 1A Operation 01 When the pressure reaches full inflation within approximately 5 minutes the normal pressure indicator will illuminate and the low pressure indicator will turn off 02 Turn the pressure dial see 5 to set a comfortable pressure level from 40 110kg according to the approximate weig...

Page 4: ...t the appliance to the nearest mains socket and ensure that the power indicator light is on Contact OSKA Service The pump appears to be on running but is not reaching the desired pressure setting The hoses are connected correctly Disconnect the hoses and re connect them in to the outlet connectors on the side of the pump Contact OSKA Service The pump is running but is not inflating the cushion The...

Page 5: ...e user owner WARNING For private non professional operators In the case of system failure please check 01 Is there sufficient power supply 02 Is the unit turned on 03 Is the power cord connected properly and is the power plug inserted correctly 04 Is the mattress connection tube connected properly If all these primary checks have been carried out and the system continues not to function then you m...

Page 6: ...Air Overlay Mattress User Guide Edward House 5 Penner Road Havant PO9 1QZ 0845 381 0600 ask oska uk com oska uk com Pressure Care Experts ...

Reviews: