Air Overlay Mattress User Guide

5

System set-up

01.   Place the pump on horizontal surface or 

hang by means of the hooks.

02.   Place the seat cushion on the chair air tubes 

should be at the rear of the chair unless 

there are no openings at the rear of the chair. 

Please ensure that the cushion is fully stable 

on the chair.

03.   Connect air feed tubes to air outlet on the 

seat cushion (see 1a) and pump (see 1b), 

ensuring that the tubes are not twisted or 

“kinked”. Allow space for tubes to drop 

freely.

04.   Switch on the pump power (see 2). 

The power switch indicator should now 

illuminate, and the pump will start to inflate 

the cushion.

05.   The low pressure indicator (see 3) will 

illuminate as inflation takes place.

06.   The audible alarm won’t alarm within 60 

minutes but will be activated if the pressure 

remains low inside the cushion. In the case of 

low pressure check the connections and the 

products. Press the alarm reset button (see 

4) to mute the audible alarm, its indicator will 

flash.

1A

Operation

01.   When the pressure reaches full inflation 

(within approximately 5 minutes) the normal 

pressure indicator will illuminate and the low 

pressure indicator will turn off.

02.   Turn the pressure dial (see 5) to set a 

comfortable pressure level from 40-110kg 

according to the approximate weight of the 

patient.

03.   Press the static button to activate static 

mode. The static indicator will illuminate. 

Static mode will commence in approximately 

5 minutes. Press the static button again to 

switch back to alternating mode.

5

4

3

1B

2

System maintenance

Maintenance at regular intervals (mandatory every 

2 years) is necessary to preserve the function of 

the OSKA SeriesC Air alternating cushion system. 
Used filters must be exchanged within this period.
• 

Replace membrane unit

• 

Replace air filter

• 

Replace synchronous motor

• 

Conduct function test/function check

• 

Sealing of case and attaching an inspector 

sticker

• 

Protective conductor check in accordance 

with DIN VDE 0751

Summary of Contents for SeriesC Air

Page 1: ...User Guide SeriesC Air ...

Page 2: ... use on a chair within home or care environment with power source nearby to ensure that comfort and relief levels are maintained The system consists of a control unit featuring a membrane pump and a cushion with 6 air cells arranged in a transverse manner The entire cell unit is encased in a cradle of visco foam The bielastic top cover and PVC outer base cover ensure stability comfort and infectio...

Page 3: ...ns and the products Press the alarm reset button see 4 to mute the audible alarm its indicator will flash 1A Operation 01 When the pressure reaches full inflation within approximately 5 minutes the normal pressure indicator will illuminate and the low pressure indicator will turn off 02 Turn the pressure dial see 5 to set a comfortable pressure level from 40 110kg according to the approximate weig...

Page 4: ...t the appliance to the nearest mains socket and ensure that the power indicator light is on Contact OSKA Service The pump appears to be on running but is not reaching the desired pressure setting The hoses are connected correctly Disconnect the hoses and re connect them in to the outlet connectors on the side of the pump Contact OSKA Service The pump is running but is not inflating the cushion The...

Page 5: ...e user owner WARNING For private non professional operators In the case of system failure please check 01 Is there sufficient power supply 02 Is the unit turned on 03 Is the power cord connected properly and is the power plug inserted correctly 04 Is the mattress connection tube connected properly If all these primary checks have been carried out and the system continues not to function then you m...

Page 6: ...Air Overlay Mattress User Guide Edward House 5 Penner Road Havant PO9 1QZ 0845 381 0600 ask oska uk com oska uk com Pressure Care Experts ...

Reviews: