Functionality of the Main Menu
52
We recommend the following two methods according to the different interface
of TSI-3000 which is selected to connect with RMM1024.
METHOD1
Use the COM7~COM12 interface of TSI-3000 to connect
with RMM1024
Cable1
Use Cable1 to connect one of the COM7~COM12 interface at the rear
panel of TSI-3000 with the RS485 Interface of RMM1024, the functionality
of the interfaces are as shown in the following table.
RS485(RMM1024): RJ45
Use the RS485 interface to connect with TSI-3000, the RS485 uses the
RJ45 connector, and the function of RS485 interface of RMM1024 is
defined as in
Table 5.1-13
:
Table 5.1-13 The Pins Input/output connectors for RS485 of RMM1024
PIN No.
RS485 IN Terminal
Signal
RS485 OUT Terminal
Signal
Illustration
1,2
GND
GND
3 Tx-
Tx-
4
Rx+
Rx+
5 Rx-
Rx-
6
Tx+
Tx+
7,8 NC
NC
COM(TSI-3000): DB9
Use one of the
COM7~COM12
interface at the rear panel of TSI-3000
to connect with RMM1024 monitor, the
COM7~COM12
has the female
DB9 connector, and the function of
COM7~COM12
interface of
TSI-3000 is defined as in
Table 5.1-14
:
Table 5.1-14 The Pins of COM* in TSI-3000
PIN No.
DB9
Illustration
1
GND
2
Rx-(data from external device)