5 TROUBLESHOOTING
TROUBLESHOOTING
The QuikRead go instrument displays error mes-
sages and guides the user in case it detects er-
rors. Follow the instruction displayed and see the
troubleshooting table in this instruction for use
Error message / Troubleshooting
Possible cause
Corrective action
QuikRead go does not start.
Power unit is not connected.
Connect power unit and retry.
The instrument has an electronic malfunction.
Contact customer service.
Touch panel on display is not functioning
properly.
Touch calibration is not correct
i.e. the active area is not under the button.
Calibrate the touch screen according to the
procedure described in the maintenance
chapter.
Touch panel is not responding at all.
Contact customer service.
Instrument alarm sounds cannot be heard.
Volume is set at a low level.
Set volume according to the procedure de-
scribed in the chapter on personal settings.
Instrument sound system has a malfunction.
Restart QuikRead go. If the problem persists,
please contact customer service.
and in the QuikRead go kit instruction for use.
Contact your local supplier in case of needed
support or a repair requirement.