26
Security settings
• Select Security login OFF or ON. When
selected ON, all users need to log on always
before starting to use QuikRead go.
• Define access to old results.
• Admin password can be changed on Man-
age the Admin account.
• Set up User and Supervisor accounts.
• Adjust the time for logout timeout.
Contact your supplier for more details about LIS
connection.
The Admin password is QRGOSET. If user ac-
counts are created, the password can be
changed.
LIS settings
From the LIS settings option, the LIS settings for
data transfer can be adjusted. The data are trans-
ferred via a serial or a LAN connection. TCP/IP
settings must be supplied before a LAN connec-
tion can function. Contact your supplier for more
details about LIS connection.
LIS features
Select the following features OFF or ON.
• LIS logging.
When set ON, LIS traffic will be
logged.
• Identify QC result.
When set ON, QC results
will have a special identification in LIS com-
munication.
• LIS delay.
When set ON, there will be a short
delay between consecutive results sent to LIS.
A delay can be useful in solving congestion
issues with older serial-type connections.
• Additional ID.
When set ON, on top of Patient
ID and Operator ID a third sample ID can be
added, e.g. patient birthdate or Doctor ID.
The additional ID can be taken into use also with
QuikRead go instruments not connected to LIS.
POCT1-A2 features
• Operator login.
When set ON, all users
need to log on always before starting to use
QuikRead go. User information has to be pro-
vided by the POCT1-A2 server.
• Patient ID validation
. When set ON, patient
ID’s will be checked against a downloaded
patient list before sending results to the LIS.
• Patient data shown
. When set ON, patient
data will be shown before a measurement
starts.
TCP/IP settings
Select the type of LIS connection used.
WLAN settings
Select the type of WLAN connection used.
Character encoding
Select the type of character encoding used for
LIS01-A2 protocol.
Contact your supplier for more details about LIS
connection.
USAGE
GMT
GMT is a universal time against which the instru-
ment’s time is set. GMT is not visible to the user
but is an internal clock for the instrument.
To set GMT adjust the date and time with the
arrow buttons.
Internal clock
• Date and time have been adjusted to GMT at
the factory.
• Date and time are stored in the memory once
a day at start up.
• If the clock battery runs out, the clock stops.
When the battery has been replaced, the
clock continues from the last saved time.
The time is adjusted in the Admin settings
and requires a password. A new adjusted time
cannot be before the last saved time. Giving
too old a time will create an error message.
Local time
• Adjust the time according to the local time
(
Settings
–
>
Measurement flow
–
>
Mainte-
nance: Date & time)
. The local time is saved
to the memory as an offset to GMT.
• When changing the clock battery, adjust the
time and date. The time cannot be adjusted
to more than 24 hours before the last saved
GMT. Giving too old a time prompts an er-
ror message ”Setting of date failed. Date is
in the past.”