18
Chapter 3
Facility Physical Requirements
Ceiling
—painted in the same manner as the walls, or covered with non-
shedding ceiling tiles.
Floors
—covered with antistatic tile, rubber, or linoleum. Carpeting is not
allowed.
Furniture
—should be resistant to static buildup. Cloth-covered seats are
preferable to plastic-ones. The feet of the furniture should
not
be insulated
from the ground, for example by rubber caps.
Yellow Room Illumination
Safe lighting should be as per specifications of the resist manufacturer.
DuPont
, for example, recommends
Encapsulite Gold 10
. Normal
illumination should also be provided for maintenance and service
procedures.