ANALOG OUTPUT / LED
The device has an alanog output corresponding
to each temperature channel. They are located
beneath each ST socket as BNC socket.
The limits are set to 0°C and 300°C.
For temperatures outside of this range analog
output is set to the maximum value.
The temperatures are mapped to values
between 0-10V or 4-20mA. If no temperature
sensor is connected, the maximum value is
applied.
The maximum values for the analog output
can be also set via the accompanying software
„FOTEMP-MKT“. For doing so, please make a
right mouse click and choose „Set Analog Out“,
for each channel.
LED
The LED indicates the status of each channel. In
case of damage, sensor either defective or non-
existence, the LED flashes red.
In normal operation, sensor is available, measu-
rements are carried out, it flashes green.