ONE POINT CALIBRATION
If a slight deviation between real and displayed
temperature is encountered, it is
possible to eliminate the temperature diffe-
rence by doing a one point calibration at a well
known temperature.
For doing so please click the right mouse
button on the temperature panel which shall
be adjusted. A context menu will appear. Please
select “one point calibration” and enter the real
temperature value and click “OK”.
The device should now display the correct
temperature.
One point calibration can be made undone at
any time selecting „reset calibration“
from the content menu.