The  performance  of  the  microscope  can’t be  made  fully  because  of  unfamiliar  using,  this  table 

will give some advices. 

Trouble 

Cause 

Remedy 

1. Optical parts 

  bulb  works,  but 

the  light  is  not 

enough 

Field diaphragm is not enough large 

Turn it larger 

The knob for brightness is too low 

Adjust to suitable 

The  polarizer  and  analyzer  is  pulled 

inside 

Take them off 

 the edge of field 

dark or not equal 

Nosepiece is not fixed 

Fix the nosepiece 

Dirt on lens or eyepiece or objective 

Clean them 

Filter,  polarizer  or  analyzer  are  not  at 

right position 

Pull  the  slides  in  or  pull  in 

the light path selection pole 

  dirt  inside  of 

field 

Dirt on specimen 

Clean specimen 

Dirt on lenses 

Clean lenses 

 image blur 

Slide cover on sample 

Take the cover away 

Sample is not upright to objective 

Adjust the sample 

Nosepiece is not fixed 

Fix the nosepiece 

Aperture diaphragm opened too large 

Adjust it 

Dirt on lens of eyepiece 

Clean it 

 

image 

blur 

somewhere 

or 

image 

moving 

during focusing 

Sample is not settled correctly 

Set it again 

Nosepiece is not fixed 

Fix the nosepiece 

  eye  tired  easily 

or double image

 

Interpupillary is not correct 

Adjust interpupillary 

Diopter is not correct 

Adjust diopter 

Not  the  same  magnification  for  both 

eyepiece 

Use same eyepiece 

 

 

                                                                        A13.0905     

 

 

Summary of Contents for A13.0905

Page 1: ...E200 series To ensure the safety and obtain optimum performance and to familiarize yourself fully with use of this microscope we recommend that you study this manual thoroughly before operating the mi...

Page 2: ...umination 11 3 2 Focusing 11 3 3 Adjust tension of focus 11 3 4 Adjust diopter 12 3 5 Adjust inter pupillary distance 12 3 6 Use filter 12 3 7 Adjust field diaphragm 13 3 8 Adjust aperture diaphragm 1...

Page 3: ...he high temperature damp dusty or acute shake place Make sure the worktable is horizontal Following environment is required Indoor temperature 5 40 Max relative humidity 80 1 3 To move the microscope...

Page 4: ...ganic solution to clean the surfaces of the microscope especially for plastic parts Please use the neutral detergent if necessary 2 3 If the microscope damped by the liquid during working please cut o...

Page 5: ...ec com 1 Nomenclature A13 0905 Mechanical stage Eyepiece Knob for adjustment tension Coarse knob Fine knob Analyser Knob for brightness Pole for field diaphragm Pole for aperture diaphragm Polarizer s...

Page 6: ...www optoedu com 6 18 sale cnoec com A13 0905 Objective IE200M stand Power switch Trinocular head Centering screw for field diaphragm Diopter adjustment ring Reflected illuminator Nosepiece Lamp holder...

Page 7: ...e below numbers shows the assemblage order Before assemblage please make sure all parts are without any dust or dirt Be careful during assemblage and do not scrap or touch the optical lens Plug Lamp h...

Page 8: ...e is finger print on bu bulb Whenever replace the bulb please cut off the power first and wait the illuminator part cool down 2 2 3 Set objectives Turn the coarse knob to make the nosepiece to lowest...

Page 9: ...d Fig 5 Turn the metal stage to make the V hole direct to the user 2 2 5 Set clips Take out clips from pole Fig 6 Screw pole into hole on mechanical stage then lock by screw nut Set the clips into pol...

Page 10: ...ator with its face up to hear click Fig 9 2 2 9 Connect power supply Make sure the main power switch is at O OFF position Insert one side of power plug into microscope socket and the other side connec...

Page 11: ...objective 5X Please make sure the specimen should be vertical with the objective If not please fix the specimen by plasticene Adjust the diopter of right tube to 0 See 3 4 Adjust diopter Observe throu...

Page 12: ...r Please make sure the diopter of standard one should be at O position before focusing 3 5 Adjust inter pupillary distance When two eyes observe please turn the two prism house with the direction indi...

Page 13: ...centering screws of field diaphragm to make the diaphragm image to the center Open the field diaphragm slowly if the field inscribed with its image it indicates the diaphragm centered Fig 18 Please i...

Page 14: ...ower switch is at 0 Off and cut off the power supply Use screwdriver to take off the fuse from fuse holder and replace a new one Fig 21 The rating for fuse 250V 8A 3 11 Mount and use TV device Loose t...

Page 15: ...ap Fig 23 Mount the photo device as the direction indicated into tri tube then lock on screw Loose the lock screw on photo adapter and take off photo tube Fig 24 Insert 3 2X photo eyepiece into tube T...

Page 16: ...quintuple nosepiece Focus Fine and coarse coaxial moving range 33mm fine precision 0 001mm Stage Size 180 155mm moving range 75 40mm 12 metal stage 25metal stage for selection Illumination Reflected...

Page 17: ...analyzer are not at right position Pull the slides in or pull in the light path selection pole dirt inside of field Dirt on specimen Clean specimen Dirt on lenses Clean lenses image blur Slide cover...

Page 18: ...ere is not power supply Connect the power supply The wires do not connect right Connect it correctly Faulty setting of bulb Set the bulb correctly Bulb burn out Replace bulb Not enough brightness Faul...

Reviews: