3.

 

Operation                                                                A13.0905     

3-1 Illumination 

  After  connect  the  microscope  with  power 

supply, set main switch at “-“ position.

 

  Turn the knob 

  for brightness to suitable. 

Clockwise turning of the adjusting knob is to 

bright and counter to dark. (Fig 11) 

 It can protect the bulb if using under low 

power supply. 

         

 

3-2 Focusing 

Put  the  specimen  on  metal  stage  and  use 

objective 5X. 

 Please make sure the specimen should be 

vertical with the objective. If not, please fix 

the specimen by plasticene

 

  Adjust  the  diopter  of  right  tube  to  “0” 

(See  3-4  Adjust  diopter).  Observe  through 

right eyepiece and turn the coarse knob 

  to 

find the image (Fig 12)

 

  Turn the fine knob 

  to make the image 

clear. 

 

 

3-3 Adjust tension of focus 

If  the  tension  is  too  tight  or  image  is  out  of 

focus  plane  or  the  stage  slips  down  during 

focusing, please adjust the tension knob 

 of 

focus. (Fig. 13) 

  The  knob 

  direction  indicated  is  for 

loose and opposite direction for tight.

 

 
 
 

 

Fig.12 

 

Fig.11 

 

Fig.13 

 

Summary of Contents for A13.0905

Page 1: ...E200 series To ensure the safety and obtain optimum performance and to familiarize yourself fully with use of this microscope we recommend that you study this manual thoroughly before operating the mi...

Page 2: ...umination 11 3 2 Focusing 11 3 3 Adjust tension of focus 11 3 4 Adjust diopter 12 3 5 Adjust inter pupillary distance 12 3 6 Use filter 12 3 7 Adjust field diaphragm 13 3 8 Adjust aperture diaphragm 1...

Page 3: ...he high temperature damp dusty or acute shake place Make sure the worktable is horizontal Following environment is required Indoor temperature 5 40 Max relative humidity 80 1 3 To move the microscope...

Page 4: ...ganic solution to clean the surfaces of the microscope especially for plastic parts Please use the neutral detergent if necessary 2 3 If the microscope damped by the liquid during working please cut o...

Page 5: ...ec com 1 Nomenclature A13 0905 Mechanical stage Eyepiece Knob for adjustment tension Coarse knob Fine knob Analyser Knob for brightness Pole for field diaphragm Pole for aperture diaphragm Polarizer s...

Page 6: ...www optoedu com 6 18 sale cnoec com A13 0905 Objective IE200M stand Power switch Trinocular head Centering screw for field diaphragm Diopter adjustment ring Reflected illuminator Nosepiece Lamp holder...

Page 7: ...e below numbers shows the assemblage order Before assemblage please make sure all parts are without any dust or dirt Be careful during assemblage and do not scrap or touch the optical lens Plug Lamp h...

Page 8: ...e is finger print on bu bulb Whenever replace the bulb please cut off the power first and wait the illuminator part cool down 2 2 3 Set objectives Turn the coarse knob to make the nosepiece to lowest...

Page 9: ...d Fig 5 Turn the metal stage to make the V hole direct to the user 2 2 5 Set clips Take out clips from pole Fig 6 Screw pole into hole on mechanical stage then lock by screw nut Set the clips into pol...

Page 10: ...ator with its face up to hear click Fig 9 2 2 9 Connect power supply Make sure the main power switch is at O OFF position Insert one side of power plug into microscope socket and the other side connec...

Page 11: ...objective 5X Please make sure the specimen should be vertical with the objective If not please fix the specimen by plasticene Adjust the diopter of right tube to 0 See 3 4 Adjust diopter Observe throu...

Page 12: ...r Please make sure the diopter of standard one should be at O position before focusing 3 5 Adjust inter pupillary distance When two eyes observe please turn the two prism house with the direction indi...

Page 13: ...centering screws of field diaphragm to make the diaphragm image to the center Open the field diaphragm slowly if the field inscribed with its image it indicates the diaphragm centered Fig 18 Please i...

Page 14: ...ower switch is at 0 Off and cut off the power supply Use screwdriver to take off the fuse from fuse holder and replace a new one Fig 21 The rating for fuse 250V 8A 3 11 Mount and use TV device Loose t...

Page 15: ...ap Fig 23 Mount the photo device as the direction indicated into tri tube then lock on screw Loose the lock screw on photo adapter and take off photo tube Fig 24 Insert 3 2X photo eyepiece into tube T...

Page 16: ...quintuple nosepiece Focus Fine and coarse coaxial moving range 33mm fine precision 0 001mm Stage Size 180 155mm moving range 75 40mm 12 metal stage 25metal stage for selection Illumination Reflected...

Page 17: ...analyzer are not at right position Pull the slides in or pull in the light path selection pole dirt inside of field Dirt on specimen Clean specimen Dirt on lenses Clean lenses image blur Slide cover...

Page 18: ...ere is not power supply Connect the power supply The wires do not connect right Connect it correctly Faulty setting of bulb Set the bulb correctly Bulb burn out Replace bulb Not enough brightness Faul...

Reviews: