background image

19 

 

 

 

As the performance of microscope can’t play fully due to unfamiliar operations, the table below 
can provide some solutions. 

Problem 

Cause 

Solution 

1. Optical Part 

 

 

(1)  The  LED  light 
is  bright,  but  it’s 
dark  in  the  field  of 
view. 

Field  diaphragm   is   not   large 
enough. 

Enlarge the field diaphragm. 

Condenser is too low. 

Adjust the condenser. 

Condenser centered incorrectly. 

Center the condenser. 

Condenser  installed  incorrectly, 
the  fixed  screw  is  not  in  the 
groove of bracket. 

 

Install it correctly. 

 

(2)  The  side  of  the 
field of view is dark 
or not even. 

The  nosepiece  is not  in  the right 
position. 

Turn  the  nosepiece  into  the  right 
position. 

Stain or dust has accumulated on 
the  lens  such  as  condenser, 
objective, or eyepiece. 

 

Clean the lens. 

(3)  Stain  or  dust  is 
observed 

in 

the 

field of view. 

Stains  have  accumulated  on  the 
specimen. 

Clean the specimen. 

Stains have accumulated on the 
lens. 

Clean the lens. 

 
 
 
 
 
 
 
 

(4) Unclear image 

The cover glass is not standard. 

Use a standard cover glass  with 
thickness 0 or 0.17mm. 

The cover glass faces down. 

Adjust it. 

The 

immersion 

oil 

has 

accumulated 

on 

the 

dry 

objective. 

 

Clean it thoroughly. 

The immersion oil is not used for 
oil objective. 

Use immersion oil. 

Air bubble is in the immersion. 

Get rid of the air bubble. 

Use wrong immersion oil. 

Use a correct one. 

The  aperture  diaphragm  is  not 
opened correctly. 

Adjust the aperture diaphragm. 

Stain or dust has accumulated on 
the inlet lens of eyepiece. 

Clean the lens. 

The condenser is too low. 

Adjust  the  condenser  to  the 
highest position. 

 

(5)  One  side  of  the 
field of view is dark 
or  the  image  moves 
while focusing. 

The nosepiece is not in the right 
position. 

Turn the nosepiece into the right 
position. 

Condenser centered incorrectly. 

Center the condenser. 

Condenser installed  incorrectly, 
the  fixed  screw  is  not  in  the 
groove of bracket. 

 

Install it correctly. 

5. Troubleshooting 

Series 

Summary of Contents for A12.0911

Page 1: ...g and maintenance to the biological microscope Please study this manual thoroughly before operating and keep it with the instrument The manufacturer reserves the rights to the modifications by technol...

Page 2: ...Adjust the Field Diaphragm 10 3 11 Aperture Diaphragm 11 3 12 Use the Oil Objective 100X 11 3 13 Use the Filter 12 3 14 Use Button and Interface of the Digital Head 12 4 Assembling and Operation of Ac...

Page 3: ...replacing the LED lamp and wait until the lamp and base cool down completely see Fig 3 Bulb selected only single 3W LED class 3B 5 External power supply adaptor is adopted Wide voltage range is suppor...

Page 4: ...on t place these chemical near to fire or fire source For example when turning on or turning off the electrical device please use these chemical in a ventilated place 2 Don t use organic solution to w...

Page 5: ...minator Light Adjustment Knob Coarse Focusing Knob Fine Focusing Knob Focus Arm Koehler Illuminator Condenser Stage Tension Adjustment Ring Objective Body Group Eyepiece Binocular Head Objective Nosep...

Page 6: ...Illuminator Condenser Koehler Illuminator Color Temperature Knob Power Switch LCD Display Coarse Focusing Knob Fine Focusing Knob Focus Arm Stage Objective Objective Nosepiece Body Group Gemel Binocul...

Page 7: ...Koehler Illuminator Condenser Focus Arm Fine Focusing Knob Koehler Illuminator LCD Display Power Switch Transmitted Color Temperature Knob Coarse Focusing Knob series biological microscope with LED fl...

Page 8: ...epiece Body Group Stage Focus Arm External Power Socket Coarse Focusing Knob Light Adjustment Knob Transmitted Fine Focusing Knob Koehler Illuminator Eyepiece Gemel Binocular Head LED Fluorescence Ill...

Page 9: ...Following is the Assembling Scheme and the numbers denote the assembling order Before assembling make sure there is no dust dirt or other materials which will disturb it Assemble carefully and do not...

Page 10: ...20X 40X 100X objective matching with the nosepiece hole 1 2 3 4 5 see Fig 6 When operating first use the low magnification objective 4X or 10X to search for specimen and focus and then replace with hi...

Page 11: ...Power Cord Don t use strong force when the power cord is bended or twisted otherwise it will be damaged 1 Make sure the power switch is at O OFF before connecting the power cord 2 Insert the connector...

Page 12: ...ording to the demand Insert a flat tip spanner into the color temperature adjustment knob to select increase or decrease with step 100 Rotate clockwise to increase color temperature and rotate counter...

Page 13: ...nob for clear details and lock the position screw The position screw can stop the objective touching the clips 3 6 Adjust the Focusing Tension If the handle is very heavy when focusing or the specimen...

Page 14: ...o raise it to the highest position see Fig 19 2 Rotate the objective 10X to the light path and focus the specimen 3 Rotate the field diaphragm adjustment ring to put the field diaphragm to the smalles...

Page 15: ...ective 3 12 Use the Oil Objective 100X 1 Use the 4X or 10X objective to focus the specimen 2 Place a drop of oil on the specimen observed see Fig 23 3 Rotate the nosepiece counterclockwise and rotate...

Page 16: ...e indicator flashes to indicate that the camera is working 2 Photo button when the camera is working press once to take a photo 3 Recording button when the camera is working the first press is the rec...

Page 17: ...nto the power socket of the microscope back and then insert the other end into the power supply socket and ensure them be connected well see Fig 28 Before connecting the external power adaptor firstly...

Page 18: ...necting the power cable firstly rotate the rechargeable battery box switch the transmission power switch and the reflection adjustment knob with switch to Off turn off 2 Turn on the battery box switch...

Page 19: ...C UL safety certification of PC 5 For binocular observation after the image is clear view the output video through the corresponding video capture analysis software If the image is unclear rotate the...

Page 20: ...ntered it doesn t need to be adjusted in operation in normal cases Fig 37 4 4 2 Assembling and Operation of Disc Phase Contrast Condenser 1 Assembling of Condenser please refer to 2 2 1 change the obj...

Page 21: ...denser flapper socket as the direction of the arrow pointed see Fig 39 2 Each diaphragm or hole has corresponding position when it sounds ka da in dark field flapper rotating it indicates that one dia...

Page 22: ...to the USB output port on the back of observation head and the other end into the USB interface of PC see Fig 41 View the microscope video by the corresponding video capture and analysis software and...

Page 23: ...lean the specimen Stains have accumulated on the lens Clean the lens 4 Unclear image The cover glass is not standard Use a standard cover glass with thickness 0 or 0 17mm The cover glass faces down Ad...

Page 24: ...e 4 Stage declines itself cannot stay on the focal plane when observing Tension knob is too loose Tighten a little 5 Coarse focusing knob can t decline The base of the condenser is too low Raise the b...

Page 25: ...field is dark The fluorescence filter block adjustment knob is at Brightfield Rotate the fluorescence filter block adjustment knob to Fluorescence 3 The background is yellowgreen when in brightfield o...

Reviews: