BWS350 System Installation, Operation & Maintenance
6
Typical Installation with 50 Gal. Atmospheric Tank
Important: Plumbing should be performed by a qualified plumber in accordance with local codes.
Feed Water Line -
Green - from Water Supply
Valve to Feed Water Inlet
Feed
Water
Optimized
(Treated)
Water to
Equipment
Feed Water
Inlet
Optimized
Water Line
- Blue - from
Optimized Water
Outlet to Tank Inlet
Drain Line
- Black -
from Reject Outlet to Drain
Reject
Water
(to Drain)
User-Supplied
Water Supply
Valve &
Pressure Gauge
Reject
Water
Outlet
Tank High Level Float Switch Cable
Optimized
Water
Outlet
to Eqpt
Optional
Carbon
Post-
Treatment
Power
120VAC or 230VAC, 2A
Optimized W
ater Line
- 1/2" Gray Hose
with hose barb insert - Make connection to distribution and (optional) post-treatment.
CT
O-Q
Pre-Filter
CT
O-Q
Pre-Filter
Emergency
Bypass Valve
Operating
Pressure
Gauge
Membrane Reject W
ater
Tubing Connection
AMS-QT
Membrane C
ar
tri
dg
e
Water Quality
Monitor
1
3
2
Processor
Tank
Repressurization
Return
Optimized
Water
to Storage
Tank
Sample
Port
Power Cord
When installing, keep lines from the Repres-
surization Assembly to the equipment as
short as possible to minimize pressure loss.
IMPORTANT - PROTECT PROCESSOR AND
REPRESSURIZATION ASSEMBLY FROM THE
ELEMENTS. DO NOT INSTALL IN DIRECT
SUNLIGHT OR WHERE EXPOSED TO FREEZ-
ING TEMPERATURES OR RAIN.
Blending Valve
-Set to req'd
TDS reading!
Repressurization
Assembly
Optimized
Water to
Storage Tank
Inlet
(on valve)
Repressurization
Assembly Outlet
(Hose Barb)
Optimized Water
Storage Tank
- 50 gal. Atmospheric
Pressurized Water Line
- 1/2" Gray Hose with hose barb inserts at both ends - from
Repressurization Assy Outlet to Storage Tank/RP Return
Safety Float
Valve
(inside tank)
Air
Breather
Repressurization
Pump
120VAC or 230VAC
Buffer
Tank
Buffer Tank Valve
(Normally Open)
Tank Inlet Divert Valve
(Valve normally in Down
position. Turn to Up position
to bypass Repressurization
Assy.)
Processor
Control High
Level Float
Switch
(inside tank)
Bypass
Low Level
Float Switch
(inside tank, with
Control Box at-
tached to Buffer
Tank bracket)
4
MA-Q15
Mi
ne
ra
l Cartridge