ASE-C-MSA
User’s Manual
Page 7 of 10
Optilab, LLC
600 E. Camelback Road, Phoenix, AZ 85012
Phone: (602) 343-1496, Fax: (602) 343-1489, Email: [email protected]
2.7
Remote Control Command Set
When the electrical connections have been made, and the software settings for
serial port transmission are set correctly, you are now able to send commands
to the ASE module. The commands are listed below:
Command
Description
Example
Response
READ
Reads the equipment basic
information.
READ
{CR,LF}
Optilab,LLC.
Model Type:ASE-C-MSA-3
SN:XXXXXXXXX
Version:VX.XX
PUMP Current:000
Protect Current:600
Output1:Low
Output2:Low
Output3:Low
SETC
Set the pump laser current.
The current can be set
between 0 mA and the
protect current (600 mA
typically).
SETC:050
{CR,LF}
Set the pump
current to be 50
mA.
PUMP Current:050
2.8
Mechanical Drawing
Unit: mm