Keratron
™
Piccolo Installation and operating manual
OPTIKON 2000
5.3.2
GUIDANCE AND MANUFACTURER’S DECLARATION –
ELECTROMAGNETIC IMMUNITY
The Keratron
™
Piccolo is intended for use in the electromagnetic environment
specified below. The customer or the user of the Keratron
™
Piccolo should
assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic Environment - Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
•
±6 kV contact
•
±8 kV air
•
±6 kV contact
•
±8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%
Electrical/fast
transient/burst
IEC 61000-4-4
•
±2 kV for power supply lines
•
±1 kV for input/output lines
•
±2 kV for power supply
lines
•
±1 kV for input/output lines
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
IEC 61000-4-5
•
±1 kV differential mode
•
± 2kV common mode
•
±1 kV differential mode
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips, short
interruptions, and
voltage variations
on power supply
input lines
IEC 61000-4-11
•
<5% U
T
(>95% dip in U
T
) for
0.5 cycle
•
40% U
T
(60% dip in U
T
) for 5
cycles
•
70% U
T
(30% dip in U
T
) for
25 cycles
•
<5% U
T
(>95% dip in U
T
) for
5 sec
•
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
•
40% U
T
(60% dip in U
T
) for
5 cycles
•
70% U
T
(30% dip in U
T
) for
25 cycles
•
<5% U
T
(>95% dip in U
T
)
for 5 sec
Mains power quality should be that of a
typical commercial or hospital environment. If
the user of the Keratron
™
Piccolo requires
continued operation during power mains
interruptions, it is recommended that the
Keratron
™
Piccolo be powered from an
uninterruptible power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
Not applicable.
The equipment does not
contain any magnetically
sensitive components.?
Power frequency magnetic field should be at
level characteristic of a typical location in a
typical commercial or hospital environment.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3V/m
80 MHz to 2.5 GHz
3 Vrms
3V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the Keratron
™
Piccolo, including
cables, than the recommended separation
distance calculated from he equation
applicable to the frequency of the transmitter.
Recommended separation distance:
d= 1.2
√
P
d= 1.2
√
P 80MHz to 800 MHz
d= 2.3
√
P 800MHz to 2.5 GHz
where P is the maximum output power rating
to the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
a, should be less than the compliance level
in each frequency range. b
Interference may occur in the vicinity of
equipment marked with the following symbol:
A