5.3 EMC TABLES
5.3.1
GUIDANCE AND MANUFACTURER’S DECLARATION –
ELECTROMAGNETIC EMISSIONS
The Keratron
™
Piccolo is intended for use in the electromagnetic environment
specified below. The customer or the user of the Keratron
™
Piccolo should assure
that it is used in such an environment.
Emissions test
Compliance
Electromagnetic Environment – Guidance
RF emissions
CISPR 11
Group 1
The Keratron
™
Piccolo uses RF energy only for
its internal function.
Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
N/A
( P <75W)
Voltage fluctuations/Flicker
emissions
IEC 61000-3-3
Complies
The Keratron
™
Piccolo is suitable for use in all
establishments including domestic establishments
and those directly connected to the public low
voltage supply network that supplies buildings
used for domestic purposes.