6212C Manual Version 15_22 Rev 07-19.docx
Page 47 of 53
Activate Failed Part Output Options
Set Values
xx = subfunction
Request Values
@07#<cr-lf>
Response
@07<cr-lf>
Set xx subfunctions
@07,xx#<cr-lf>
00#
Turn off failed part output.
01#
Any value above max roughness or below min roughness encountered between start and stop,
contact closes after stop received
02#
All values above max roughness or below min roughness encountered during start and stop,
contact closes and opens with bad and good values
03#
Average value above max roughness or below min roughness encountered between start and
stop, contact closes after stop received
@10#<cr-lf> Laser on Voltages: used to request just laser on voltages. Only for testing purposes
Response:
@10<cr-lf>
0.123456<cr-lf>
Detector 1 volts
0.123456<cr-lf>
Detector 2 volts
Same for Detectors 3-37
Detector 3-37 volts
12.3456<cr-lf>
Sum Voltage of Detectors 1-37
#<cr-lf>
@11#<cr-lf> Convert Light & subtract Dark, correct and send processed voltages
Note:
Responses should be background subtracted and have gain resistor correction
Responses should NOT have any other corrections
Response:
@11<cr-lf>
0.123456<cr-lf>
Detector 1 volts
0.123456<cr-lf>
Detector 2 volts
Same for Detectors 3-37
Detector 3-37 volts
12.3456<cr-lf>
Sum Voltage of Detectors 1-37
#<cr-lf>