Individual Parts / Parts Inspection
Figure 3
–
Lasercheck Head Mounted on Conveyor for Automated Inspection
There are ten drilled and tapped holes on the Lasercheck sensor head that can be used for mounting and
installing the Lasercheck in an automated inspection application. The head should be positioned at a location
where parts will pass beneath the gage at the correct vertical and horizontal position relative to the gage head
(see appendix section on Lasercheck Alignment Principles and Procedures). If parts are not clean, an air knife
should be installed and mounted prior to the laser head to blow-dry excess coolant off of the surface to be
measured. Ideally the parts would pass a few millimeters under the air knife, which would operate with a
pressure of approximately 20 psi blowing on the surface.
Positioning Prox Sensors on Conveyor for Automated Inspection
Prox Sensor Positioned to Start as Parts Enter and Leave Measurement Position
Lasercheck Sensor
Lasercheck Sensor Mounted on Conveyor for Automated Single Measure Inspection
“Proxy Sensors
Air Knife
Parts
Parts Conveyor
Start Proxy Sensor
Lasercheck Sensor
Parts
Proximity Sensor Trigger positioned to activate as surface reaches Measurement Position