LBL-2019-41-v4 OptoMonitor 3 IFU wTAVI_en Rev 2022-03
34
|
P a g e
-
A dPR pullback recording will display the recorded measurements and the dPR pullback value
over time in the graphical window. Minimum dPR over the pullback will display in the data
window, as well as be marked by a purple line at the point of minimum dPR in the graphical
window. No values for Pa, Pd, or Pd/Pa will be displayed in this playback mode. Pressing the dPR
button in the data window will return the marker to the point of minimum dPR.
-
Pressing the Pd/Pa button in the information window while in a dPR pullback playback will cause
the playback to switch to “Pd/Pa” mode, where the graphical window will be updated with Pd/Pa
over time and be marked by a yellow line at the point of minimum Pd/Pa, as well as update the
data window accordingly. Pressing the Pd/Pa button in the data window will return the marker
to the point of minimum Pd/Pa.
Pd/Pa Button
dPR Button