The power supply to the Display Unit has one LED, which will have a steady
Green
to indicate that the DU
is powered
ON
.
3.6.2.2
General system status
OU power: Flashing blue
System not ready to use. Refer to messages in the Display
Unit’s Message window.
OU top row LEDs: Third flashing, fourth
solid
The OU is updating. Refer to section 8.9 to update OU.
OU top row LEDs: All flashing
System is in error. Restart the OU. If the LEDs are still flashing
after restarting the OU, please contact OpSens or personnel
authorized by OpSens.
3.6.2.3
Handle Unit Indicators:
The Handle Unit green and amber LED indicators provide information using color and flash-coded
sequences. Please refer to table below for more details.
The flash-coded sequences are represented graphically on the Handle unit as follow:
Fast pulsing
Slow pulsing (every 2 – 3 seconds)
Slow flashing
Steady on
NOTE: Amber LED indications usually mean the system requires operator attention.
Events
Led indicators
Message on Display Unit
GFC connector not inserted to
Handle Unit – Insert GFC
Connect GF connector to Handle Unit
Optical connection loss –
Connect/Check OW
No signal, check OptoWire connection
No signal, check SavvyWire connection
System ready/Zero
completed
Wire Zero failed/Aborted
OptoWire zero aborted (signal unstable)
OptoWire zero aborted (signal unstable),
, old value used
SavvyWire zero aborted (signal unstable)
GFC & OW mismatch
Mismatch between OptoWire and GF
connector