The following registers contain the display units strings:
Registers
Variable
Data Type
Units
Function Code
Addresses
32007-32012
Volume flow units
string
—
03, 04
2006-2011
32001-32006
Mass flow units
string
—
03, 04
2000-2005
32025-32030
Temperature units
string
—
03, 04
2024-2029
32019-32024
Pressure units
string
—
03, 04
2018-2023
32031-32036
Density units
string
—
03, 04
2030-2035
32013-32017
Energy flow units
string
—
03, 04
2012-2017
Function codes 03 (read holding registers) and 04 (read input registers) are the only codes
supported for reading these registers, and function codes for writing holding registers are not
implemented. We recommend that the floating point and long integer registers be read in a single
operation with the number of registers being a multiple of two. If this data is read in two separate
operations, each reading a single 16-bit register, then the value will likely be invalid.
The floating point registers with values in display units are scaled to the same units as are
displayed, but are instantaneous values that are not smoothed. If display smoothing is enabled
(non-zero value entered in the Display TC item in the Display Menu), then the register values
will not agree exactly with the displayed values.
Exception Status Definitions
The Read Exception Status command (function code 07) returns the exception status byte,
which is defined as follows. This byte may be cleared by setting “coil” register #00003 (function
code 5, address 2, data = 0xff00).
Bit(s)
Definition
0-1
Byte order (see Modbus Order on page 2)
0 = 3-2:1-0 1 = 2-3:0-1
2 = 1-0:3-2 3 = 0-1:2-3
2
Temperature sensor fault
3
Pressure sensor fault
4
A/D converter fault
5
Period overflow
6
Pulse overflow
7
Configuration changed
Discrete Input Definitions
The status of the three alarms may be monitored via the Modbus Read Discrete Input command
(function code 02). The value returned indicates the state of the alarm, and will be #1 only if
the alarm is enabled and active. A zero value is transmitted for alarms that are either disabled or
inactive.
Registers
Variable
Function Code
Address
10001
Alarm #1 state
02
0
10002
Alarm #2 state
02
1
10003
Alarm #3 state
02
2