F-1500 Turbine Flow Meter Manual 02/18 - 1018-6 / 35726
Page 48
3.3.12 Calibration Menu
If the meter display indicates a temperature or pressure fault, a substitute value can be entered
to allow flow calculations to continue at a fixed value until the source of the fault is identified
and corrected. The units of measure of the displayed values are the same as the units
configured for the flow meter.
Password
ENTER
Use
keys to access menus.
Calibration
Menu
Meter Size
or Pipe ID
ENTER
Run Mode
Meter Factor
xxxx
Vortex Coef Ck
xx
Low Flow Cutoff
xx
Serial Number
xxxxxxxxx
F-2700 Series – meter size
F-2600 Series – pipe internal diameter (inches)
Meter calibration constant
– pulses / ft
3
– pulses / ft
Adaptive filter setting
Low Flow Cutoff
setting displayed
in volumetric flow
units (view only)
< Vol (xxx/xxx) >
xxx
< Mass (xxx/xxx)
xxx
Low Flow Cutoff
setting displayed
in mass flow
units (view only)
F-2700 Series
F-2600 Series
The Calibration Menu contains the calibration coefficients for the flow meter. These
values should be changed only by properly trained personnel. The Coef Ck and Low Flow
Cutoff are set at the factory. Consult the factory for help with these settings if the meter is
showing erratic flow rate.
3.3.13 Password Menu
Password
ENTER
Use
keys to access menus.
Password
Menu
Set Password
1234
ENTER
Run Mode
Use the Password Menu to set or change the system password. The factory-set password is 1234.