OMS ARCADIA EXT S.T.01/3R ED.0  REV.8  09/2020  

55 

 

 

7.

 

TECHNICAL INSTRUCTIONS FOR INSTALLATION 

7.1.

 

PACKING 

The unit shipped enclosed in a box (weight approx. 165 kg), contains: 

Patient chair, fixing screws patient chair, right armrest *, suction *, operating instructions, wiring diagrams, 
guarantee coupon  

(*)Present only in the case of a specific request to the order. 

Outside each case there is a mark showing the order number, the order confirmation number, the device 
model. 

N.B.: The packing of a possible seat O.M.S. is separate 

MOUNTING   

1.

 

first operation to do, the Chair must be unpacked and placed in the vicinity of the predispositions 

previously made using the installation plan in possession of the local organisation. 

2.

 

Attach the seat to the chair, insert the headrest and right armrest. 

3.

 

Record any "games" by screwing or unscrewing the adjustment feet (supplied). 

4.

 

Attach the chair to the floor using the two expansion screws (supplied). 

5.

 

Make the electrical connection and proceed with the installation of the cover. 

6.

 

Connect the ground wires to the screw marked with the earth symbol. 

7.

 

Insert the headrest and right armrest. 

7.2.

 

TESTING AND IMPLEMENTATION 

1.

 

. Check the correct operation of the manual descent and ascent pantograph and backrest controls, the 

automatic reset control, the rinse position and the "last-position" working position. Memorize and 
verify the good positioning of the 3 programs (see Manual controls and 4.3 automatic controls). 

2.

 

Check the correct operation of the limit switches and safety systems (see Safety Systems Section).

 

 

IMPORTANT WARNING: 

During the testing, it is necessary to ensure the mechanical stability of the device, after having 
placed all the movable elements and accessories (tablet, lamp, water group, etc.) in the most 
unfavourable position. If it is not possible to fix the chair to the floor, stabilizer elements are 
available. 

 

 

DANGER 

Summary of Contents for ARCADIA EXT

Page 1: ...Instructions for use Patient Chair ARCADIA EXT...

Page 2: ...PLATE 19 3 COMMAND CONTROL WARNING 20 3 1 SAFETY SYSTEMS 20 3 2 CHAIR COMMAND FROM FOOTCONTROL 21 3 3 MANUAL COMMAND FROM CHAIR 23 3 4 AUTOMATIC COMMANDS FROM THE CHAIR 24 4 CHAIR 30 4 1 DOUBLE JOINT...

Page 3: ...COMPATIBILITY IN ACCORDANCE WITH IEC 600601 1 2 57 9 DISPOSING OF THE DEVICE AT THE END OF LIFE 61 10 REPORTING ACCIDENTS TO PEOPLE 62 ATTENTION Before using the device it s necessary to read carefull...

Page 4: ...lly and uniform loads of the patient s body whatever its size The backrest can have a counterbalancing entry optional which prevents the patient the uncomfortable pulling the clothes during the descen...

Page 5: ...ituation which may cause a direct connection to serious injury or death NOTE Warning explanation or integration important information for users and technicians 1 2 SERVICE O M S SPA Officine Meccanich...

Page 6: ...d in spaces with electrical systems in accordance with the regulations in force in your country DANGER The installation of the device must be done from an authorized O M S technician the choice of pip...

Page 7: ...OMS ARCADIA EXT S T 01 3R ED 0 REV 8 09 2020 7 Replace the faulty or worn parts only with original spare parts guaranteed O M S...

Page 8: ...on approved accessories and or unauthorized modification can constitute imminent danger of injury to persons and damage to materials ATTENTION All maintenance operations must be performed after switch...

Page 9: ...RCADIA EXT S T 01 3R ED 0 REV 8 09 2020 9 The main switch achieves the isolation of the unit from the mains direct power supply before making any operation inside the unit make sure that the switch is...

Page 10: ...1 2 medical electrical equipment collateral regulation electromagnetic compatibility DANGER In the version with ground water box during the ascent of the patient chair check the patient positioning on...

Page 11: ...OMS ARCADIA EXT S T 01 3R ED 0 REV 8 09 2020 11 Do not sit at the end of the head or feet of the chair...

Page 12: ...s before effecting any moving of the platform of the patient chair detach power supply the power supply of the patient chair must be connected only and only after the correct positioning with the colu...

Page 13: ...RCADIA EXT S T 01 3R ED 0 REV 8 09 2020 13 Patient chair on wheels the positioning of the chair should be done without the patient ATTENTION Patient chair on wheels handling area must be completely cl...

Page 14: ...NTY The product is covered by a warranty period of 36 thirty six months from the date of installation from the purchaser The warranty conditions are shown inside the warranty book supplied with the de...

Page 15: ...evel device IPX0 Protection level foot control IPX1 POWER SUPPLY Rated voltage 230 Vac 10 Rated frequency 50 60 Hz Network connections that comply with the rules in force in the territory Rated curren...

Page 16: ...R ED 0 REV 8 09 2020 16 Maximum load permitted on the patient chair 200 Kg with cupidor on the floor not attached 155Kg with cuspidor attached to chair Minimum space recommended for installation about...

Page 17: ...MS ARCADIA EXT S T 01 3R ED 0 REV 8 09 2020 17 DIMENSIONS 410 850 NOTE The measures are to be considered indicative and might change according to the configuration of the dental unit and backrest type...

Page 18: ...for basic safety and essential performance of dental equipment ISO 6875 patient chairs UNI CEI EN ISO 14971 risk analysis CEI EN 62304 ing Software CEI EN 62366 ing Usability CEI EN 60601 1 6 Usabilit...

Page 19: ...that brings the general data of the instrument Device model Rated voltage and nature of current alternating Rated frequency Rated current Rated power Device serial n Manufacturer WEEE symbol Applied...

Page 20: ...limiting the ascent of the chair at a defined height DANGER In the version with a floor water group during the ascent of the chair check the positioning patient on the arm to prevent the risk of crush...

Page 21: ...HAIR COMMAND FROM FOOTCONTROL The footcontrol has two side levers B and C that allow the operator to position the chair using the controls on the foot Ascent command pantograph Descent command pantogr...

Page 22: ...of the chair are equipped with electro mechanical limit switches Once you have reached the extremes of the pantograph or backrest they automatically break the power of the engine in question thus avoi...

Page 23: ...ands on the upper end of the backrest of the Chair see figures 3 and 4 These are the functions of each A Ascent backrest command push the button towards the outside B Descent backrest command push the...

Page 24: ...THE DENTAL UNITS WITH SOSPEND WATER GROUP LINEA ESSE LINEA PATAVIUM TEMPO 9 ELX 3 4 1 1 STANDARD VERSION The operation of the controls is as follows E Rinse command push the button towards the outsid...

Page 25: ...OMS ARCADIA EXT S T 01 3R ED 0 REV 8 09 2020 25...

Page 26: ...eset this command activates the descent of the patient chair pantograph in the minimum position of the backrest simultaneously up to the maximum of its stroke When the Chair is in the zero position pr...

Page 27: ...other than the one reset this command activates the descent of the patient chair pantograph in the minimum position and simultaneously the backrest up to the maximum of its travel When the chair is to...

Page 28: ...memory recall 1 push the button towards the inside starting from any position other than the one reset this command activates the descent of the patient chair pantograph in the minimum position and si...

Page 29: ...ards the inside starting from any position other than that of zeroing this command activates the descent pantograph of the chair in a minimum position and at the same time the ascent of the backrest t...

Page 30: ...the back It makes the headrest it easy to move on two rotation axes By carrying the headrest cushion to an inclination of about 90 with respect to the swivel mechanism it can be rotated on itself by...

Page 31: ...OMS ARCADIA EXT S T 01 3R ED 0 REV 8 09 2020 31...

Page 32: ...the rear it makes it easy to move the headrest on two rotation axes By carrying the headboard cushion to an inclination of about 90 with respect to the swivel mechanism it can be rotated on itself by...

Page 33: ...OMS ARCADIA EXT S T 01 3R ED 0 REV 8 09 2020 33...

Page 34: ...ith an excursion of about 23 cm and has a articulation on the rear joint that also allows the lateral movement using the lever placed in the rear It makes it easy to move the headboard on three rotati...

Page 35: ...OMS ARCADIA EXT S T 01 3R ED 0 REV 8 09 2020 35...

Page 36: ...t forward or backward to facilitate the passage of the patient Once the armrest is pulled back it is also possible to remove it simply by pulling it in the direction of the axis of rotation helping wi...

Page 37: ...rect positioning with the column of the dental unit ATTENTION Patient chair on wheels disconnect the connection between the chair and the unit before carrying out any movement of the footrestchair DAN...

Page 38: ...uence shown below for a correct position of the patient chair 1 Press the release pedal to move the chair and close to the column of the unit on the side of the Spittoon Detail at neutral pedal press...

Page 39: ...OMS ARCADIA EXT S T 01 3R ED 0 REV 8 09 2020 39 2 Approximate with the left side of the base of the chair parallel to the base of the column...

Page 40: ...OMS ARCADIA EXT S T 01 3R ED 0 REV 8 09 2020 40...

Page 41: ...atient chair side by side with the unit so that the bases on the short edge are aligned in the rear and so that the bases on the long side are perfectly parallel The plate protruding to the left of th...

Page 42: ...OMS ARCADIA EXT S T 01 3R ED 0 REV 8 09 2020 42 The plate protruding to the left of the base of the chair must be positioned as in the picture This gives maximum mechanical stability...

Page 43: ...n the ground using the pedal DANGER Patient chair on wheels before sitting the patient on the patient chair operate the pedal for the ground lock Detail pedal with lowered wheel press the pedal on the...

Page 44: ...OMS ARCADIA EXT S T 01 3R ED 0 REV 8 09 2020 44...

Page 45: ...OMS ARCADIA EXT S T 01 3R ED 0 REV 8 09 2020 45 5 Attach the connector to the base of the column 6 Feed the chair using the appropriate power plug ELECTRICK SHOCK DANGER...

Page 46: ...Patient chair on wheels before carrying out any movement of the footrest of the chair disconnect the main plugs the electric power plug of the chair must be connected only and only after the correct...

Page 47: ...classified in Cleaning disinfecting adjustments scheduled maintenance extraordinary maintenance View all the configurations using the various accessories available in the list and therefore must be ap...

Page 48: ...ient chair For the thorough cleaning of the upholstery of the patient chair you can use a product suitable for the treatment of skay such as EMULSIO For daily cleaning and disinfection use a little ag...

Page 49: ...ly cleaning and disinfection must be done with the use of unaggressive products for the upholstery in the Skay of the patient chair The periodical thorough cleaning of the upholstery armrests and head...

Page 50: ...fety ATTENTION It is necessary to perform periodic safety tests and checks at least every two years in order to identify any electrical failures e g damaged insulation which may reduce the initial saf...

Page 51: ...OMS ARCADIA EXT S T 01 3R ED 0 REV 8 09 2020 51 The use of the dental patient chair is permitted only if all safety checks have been successful past...

Page 52: ...Who makes it Notes References Paint retouching When necessary Trained operator Fuse Replacement When necessary Trained operator Refer to the technical documentation Interventions for any malfunction...

Page 53: ...evice On the base of the armchair the protection fuses B and C can easily be identified that can be replaced externally by means of a screwdriver A series of spare fuses provided by O M S Is located i...

Page 54: ...e protective cover can be removed by unscrewing the screws D and G the cover F covering the screw G will simply be removed by pulling it out Chair board F1 10AT 32 Vdc A series of spare fuses supplied...

Page 55: ...djustment feet supplied 4 Attach the chair to the floor using the two expansion screws supplied 5 Make the electrical connection and proceed with the installation of the cover 6 Connect the ground wir...

Page 56: ...DIA EXT S T 01 3R ED 0 REV 8 09 2020 56 The main switch realizes the insulation of the appliance from the direct power supply before making any operation inside the appliance make sure that the switch...

Page 57: ...sion test Complience Electromagnetic Environment Guide Emission RF CISPR 11 Gruop 1 The device uses RF energy only for its internal operation As a result its RF emissions are very low and probably doe...

Page 58: ...etic field at Network frequency 50 60 Hz IEC 61000 4 8 3 A m 3 A m The magnetic fields at network frequency should have characteristic levels of a typical locality in a environment such as commercial...

Page 59: ...eoretically with precision An electromagnetic survey of the site should be considered to evaluate an electromagnetic environment caused by fixed RF transmitters If the field strength measured in the p...

Page 60: ...to the frequency of the transmitter where p is the maximum rated output power of the transmitter in Watts W according to the manufacturer of the transmitter NOTE 1 to 80 MHz and 800 MHz applies the se...

Page 61: ...must therefore confer the equipment reached at the end of the life of the appropriate waste collection centres of electronic and electrotechnical or return it to the retailer at the time of purchase...

Page 62: ...__________________________________________ _______________________________________________________________________________________________ Proposed operational decisions_______________________________...

Page 63: ......

Reviews: