OMS ARCADIA EXT S.T.01/3R ED.0 REV.8 09/2020
29
3.4.4.
ARCADIA EXT COMBINE TO DENTAL UNITS WITH FLOOR WATER GROUP:
VIRTUOSUS CLASSIC, CARVING PLUS AND UNIVERSAL ML
The operation of the controls is as follows:
E.
Rinse command (push the button towards the outside): Starting from any position, this command
activates the ascent of the backrest and the seat to allow the patient to access the rinsing glass;
F.
Reset command/ memory recall 1 (push the button towards the inside): starting from any position
other than that of zeroing, this command activates the descent pantograph of the chair in a minimum
position and at the same time the ascent of the backrest to the maximum of its stroke; When the
patient chair is in the zeroing position, pressing the button activates the recall of memory 1 of the
chair.
G.
Program selection in instrument table
H.
Not available
PROGRAM STORAGE
To store the channel position simply manually move the chair to the desired position (using the manual
controls) and then press and hold the button for the program that you want to memorize (F, G or H) until you
hear the beep confirming.