RUN Mode
Section 4-3
116
4-3-4
Zero Reset Function
Measured values for the workpiece after the zero display is reset can also
be displayed as negative values in reference to the new zero point that is
set. The judgement outputs will be based on the display values. The zero
reset function is effective when judging tolerances for workpieces.
n
Procedure
Set the mode switch to RUN.
Press the ENT Key for about 1 second
or longer without executing the zero
reset.
The zero reset can also be performed
using the external zero reset input. The
operation can be repeated as required.
RUN T FUN
The main display will be filled with zeros
and the zero reset indicator will light.
The linear output will be the center
value between the two points that are
set for the monitor focus.
Defaults: 0 V, 12 mA
Note 1: Maximum display range on the
negative side is
−
19999 after zero
reset.
Note 2: When the zero reset memory
function is enabled (it is enabled by
default), the zero reset value will be
stored.
Note 3: Turn OFF the zero reset mem-
ory function when the zero reset is used
for every workpiece judgement.
Press the ENT Key.