background image

12

Ⅻ.  MAINTENANCE INSTRUCTIONS

Warning: Before maintaining the oxygen concentrator, firstly cut off the 

power supply to avoid electric shock. Do not disassemble the device.
1. Outer shell

The outside of the device should be cleaned and disinfected at least once a month. If there is 
pollution on the outside , please clean and disinfect immediately. Before cleaning the oxygen concentra-
tor, you must turn off the power of the device and disconnect the AC power. Wipe the surface of the 
oxygen concentrator and the cable with a clean, soft, slightly damp, lint-free cloth or sponge. If necessary, 
wipe it with a small amount of non-corrosive diluted detergent. Wipe the device dry with a clean, dry, soft 
cloth, and do not allow liquid to penetrate into the cracks of the shell.
During the cleaning process, you only need to wipe the outer periphery of the connecting socket, not 
the inside. The cleaning agent should be removed after cleaning. Non-corrosive cleaners should be 
used to clean the outer surface of the device and the display screen. Most cleaners must be diluted 
before use.

2. Filters

The cleaning and the replacement of the filters protect the compressor and molecular sieve, and 

Symbol                                         State

Status 

Indicator

Alarm 

sound

Compressor failure alarm:
1. The compressor circuit is open (display E1) 
2. The Compressor short circuit (display E2) 

50%≤oxygen purity<82%

Power failure alarm

None

Long ring

Low priority

Run

Low priority Shutdown

Not 

applicable

≥55db(A)

≥55db(A)

Red 

light on

Long ring

Low priority Shutdown

Red 

light on

Yellow 

light on

Alarm 

sound 

pressure

Alarm 

status 

priority

State

11. Alarm and hint:

◆ After the oxygen concentrator has been running for 5 minutes, the oxygen purity monitor starts to 
work normally.
◆ After an alarm occurs, you can turn off the alarm by pressing the "ON/OFF" button of the oxygen 
concentrator.
◆ When the oxygen purity is less than 50%, the product will give out a low oxygen purity prompt, the 
display shows "E4", the red light is on, and the sound indicates.
◆ Under normal conditions, if the product has a low oxygen purity alarm or prompt, please turn it off 
immediately, use spare oxygen, and immediately contact the OMRON customer support for 
consultation (see the back cover). 

extend the oxygen concentrator life. Please clean or replace in time.

Primary filter: 

Remove the cover of the primary filter, take out the filter sponge to clean it once 

every half a month. Use light detergent firstly, rinse with clean water, and then install it after it is 
completely dry.

Secondary filter: 

After 3000 hours of use, or the filter becomes dirty, please replace it with a new 

one. Replacement method: Unscrew the fixing screws of the secondary filter cover, remove the cover, 
pull out the old filter, replace a new one, and reinstall the cover.
For consulting about filter, please contact customer support (see the back cover).

Summary of Contents for Y-506W

Page 1: ...manual Thank you for purchasing the OMRON oxygen concentrator In order to use the device safely please read the complete manual carefully before using it for the rst time Keep this instruction manual...

Page 2: ......

Page 3: ...scope of application of the product Product features Know your device Operation steps Maintenance instructions XIII Schematic diagram of gas circuit and electric circuit XIV Troubleshooting XV Compone...

Page 4: ...customer shall not disassemble and repair it privately If abnormal phenomena such as the alarm is found please contact the customer support for consultation see the back cover SAFETY INSTRUCTIONS 1 Op...

Page 5: ...ou begin to feel ill or are experiencing discomfort while using the oxygen concentrator please go to the doctor immediately 16 When the user is a child or a person with limited mobility the device nee...

Page 6: ...e stored in an environment like this No strong sunlight no corrosive odor well ventilated and clean Avoid violent vibration inverted or horizontal placement during transportation THE SCOPE OF APPLICAT...

Page 7: ...tor oxygen purity better and understand the use condition of patients oxygen purity 82 only the green light of the operation indicator is on 50 oxygen purity 82 yellow light is on oxygen purity 50 the...

Page 8: ...ce should not be used with ammable anesthetic gas mixed with air or ammable anesthetic gas mixed with oxygen or nitrous oxide 5 years except fragile parts and consumables Maximum outlet pressure Compr...

Page 9: ...omizer Primary ltration 06 Rear view of the whole device Side view of the whole device Power switch Including fuse Nebulization outlet Secondary lter Humidi cation bottle Device power jack Primary lte...

Page 10: ...he device will stop oxygen producing and enter the standby state after the timing is completed Set the timing for 8 hours and press the Timing button again to enter the continu ous oxygen therapy mode...

Page 11: ...oxygen outlet of the humidi cation bottle was blocked the pressure relief valve of the humidi cation bottle will operate by itself At this time oxygen will be ejected from the pressure relief valve bu...

Page 12: ...the network power socket and the power cord connector is inserted into the device power jack Then the device can get power supply 9 Primary lter Contains lter sponge for air ltration 10 Secondary lte...

Page 13: ...cation bottle and then put the nasal oxygen cannula on the user s ears inserting the nasal congestion in the user s nose as shown in the gures below Please follow the doctor s advice for the time and...

Page 14: ...zation kit 2 When the nebulization function is used the oxygen purity will be reduced and the device may alarm and the alarm will be automatically released when the oxygen ow is reduced It is not a ma...

Page 15: ...the device and the display screen Most cleaners must be diluted before use 2 Filters The cleaning and the replacement of the lters protect the compressor and molecular sieve and Symbol State Status In...

Page 16: ...sulting about lter please contact customer support see the back cover 3 Humidi cation bottle The oxygen concentrator is equipped with a humidi cation bottle and it is recommended to clean it each time...

Page 17: ...new fuse Figure 3 reinstall the fuse holder Figure 4 Figure 4 Figure 3 Figure 2 Figure 1 Air Filter Compressor Oxygen and nitrogen separation Adsorption tower Humidi cation bottle Oxygen outlet Oxyge...

Page 18: ...ss the ON OFF button of the operation panel as required 3 Turn on the power switch and restart the device 4 Replace the fuse Firstly cut o the power supply and carry out by professional maintenance pe...

Page 19: ...3 If the phenomenon continues please contact the customer support for consultation see the back cover 7 The inside of the nasal oxygen cannula has a lot of fog or water droplets the method of removing...

Page 20: ...it has been deformed during cleaning and disinfection 2 Replace the disposable kit 3 The cup mouth of the nebulization cup is blocked 3 Clean or replace the nebulization cup 4 Excessive addition of li...

Page 21: ...with the corresponding national laws and regulations It is recommended to consult the environmental protection department or the Omron customer support when dealing with the corresponding waste and r...

Page 22: ...n the manual 2 Portable and mobile radio frequency communication equipment may a ect the normal operation of this product 3 The maximum length of the power cord of this product is 2m 4 In addition to...

Page 23: ...le for use in all establishments including domestic establishments and those directly connected to the public low voltage network that supplies buildings used for domestic purposes Harmonic Emissions...

Page 24: ...the test voltage is applied Floors should be wood concrete or ceramic tile If oors are covered with synthetic material the relative humidity should be at least 30 Mains power quality should be that o...

Page 25: ...he eld strengths should be less than 3V m Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distanc...

Page 26: ...e by maintaining a minimum distance between portable and mobile RF communications equipment trans mitters and this concentrator as recommended below according to the maximum output power of the commun...

Page 27: ......

Page 28: ...he customer support for consultation 5 When performing warranty service if necessary we can request quali ed technicians identi ed by us to provide product circuit diagrams and repairable component da...

Reviews: