Table 17.2

Guidelines and Manufacturer's Declaration-Electromagnetic Immunity

The concentrator is intended for use in the electromagnetic environment specified below. The user of the 

concentrator should make sure it is used in such an environment.

Immunity Test

Electromagnetic Environment-

-Guidance

Electrostatic discharge 

(ESD)

IEC 61000-4-2

±6kV contact
±

8kV air

±6kV contact
±8kV air

Electrical fast transient 

burst IEC 61000-4-4

±2kV for power supply 

lines
±1kV  for input/output 

lines    

±2kV for power supply 

lines
±1kV  for input/output 

lines    

Surge IEC 61000-4-5

±1 kV line(s) to line(s)

±2 kV line(s) to earth

±1 kV line(s) to line(s)

±2 kV line(s) to earth

Voltage dips, short 

interruptions and 

voltage variations on 

power supply input 

lines IEC 61000-4-11

Power frequency

magnetic field

(50Hz/60Hz

EC61000-4-8

3A/m

3A/m

Note: U

T

 refers to the AC network voltage before the test voltage is applied.

Floors should be wood. concrete 

or ceramic tile. If floors are 

covered with synthetic material, 

the relative humidity should be 

at least 30%.

Mains power quality should be 

that of a typical commercial

or hospital environment.

Mains power quality should be 

that of a typical commercial

or hospital environment.

Mains power quality should be 
that of a typical commercial or 
hospital environment. If the user 
of the [ME EQUIPMENT or ME 
SYSTEM] requires continued 
operation during power mains 
interruptions, it is recommended 
that the [ME EQUIPMENT or ME 
SYSTEM] be powered from an 
uninterrupted power supply or a 
battery.

<

5%  U

(>95% dip in 

U

T

)for 0.5 cycle

40% U

(60% dip in 

U

T

)for 5 cycles

70% U

(30% dip in 

U

T

)for 25 cycles

<

5% U

(>95% dip in 

U

T

)for 5 sec

<

5%  U

(>95% dip in 

U

T

)for 0.5 cycle

40% U

(60% dip in 

U

T

)for 5 cycles

70% U

(30% dip in 

U

T

)for 25 cycles

<

5% U

(>95% dip in 

U

T

)for 5 sec

Power frequency magnetic 

fields should be at levels

characteristic of a typical 

location in a typical hospital or 

home environment.

IEC 60601

Test Level

Compliance 

Level

Summary of Contents for Y-506W

Page 1: ...manual Thank you for purchasing the OMRON oxygen concentrator In order to use the device safely please read the complete manual carefully before using it for the rst time Keep this instruction manual...

Page 2: ......

Page 3: ...scope of application of the product Product features Know your device Operation steps Maintenance instructions XIII Schematic diagram of gas circuit and electric circuit XIV Troubleshooting XV Compone...

Page 4: ...customer shall not disassemble and repair it privately If abnormal phenomena such as the alarm is found please contact the customer support for consultation see the back cover SAFETY INSTRUCTIONS 1 Op...

Page 5: ...ou begin to feel ill or are experiencing discomfort while using the oxygen concentrator please go to the doctor immediately 16 When the user is a child or a person with limited mobility the device nee...

Page 6: ...e stored in an environment like this No strong sunlight no corrosive odor well ventilated and clean Avoid violent vibration inverted or horizontal placement during transportation THE SCOPE OF APPLICAT...

Page 7: ...tor oxygen purity better and understand the use condition of patients oxygen purity 82 only the green light of the operation indicator is on 50 oxygen purity 82 yellow light is on oxygen purity 50 the...

Page 8: ...ce should not be used with ammable anesthetic gas mixed with air or ammable anesthetic gas mixed with oxygen or nitrous oxide 5 years except fragile parts and consumables Maximum outlet pressure Compr...

Page 9: ...omizer Primary ltration 06 Rear view of the whole device Side view of the whole device Power switch Including fuse Nebulization outlet Secondary lter Humidi cation bottle Device power jack Primary lte...

Page 10: ...he device will stop oxygen producing and enter the standby state after the timing is completed Set the timing for 8 hours and press the Timing button again to enter the continu ous oxygen therapy mode...

Page 11: ...oxygen outlet of the humidi cation bottle was blocked the pressure relief valve of the humidi cation bottle will operate by itself At this time oxygen will be ejected from the pressure relief valve bu...

Page 12: ...the network power socket and the power cord connector is inserted into the device power jack Then the device can get power supply 9 Primary lter Contains lter sponge for air ltration 10 Secondary lte...

Page 13: ...cation bottle and then put the nasal oxygen cannula on the user s ears inserting the nasal congestion in the user s nose as shown in the gures below Please follow the doctor s advice for the time and...

Page 14: ...zation kit 2 When the nebulization function is used the oxygen purity will be reduced and the device may alarm and the alarm will be automatically released when the oxygen ow is reduced It is not a ma...

Page 15: ...the device and the display screen Most cleaners must be diluted before use 2 Filters The cleaning and the replacement of the lters protect the compressor and molecular sieve and Symbol State Status In...

Page 16: ...sulting about lter please contact customer support see the back cover 3 Humidi cation bottle The oxygen concentrator is equipped with a humidi cation bottle and it is recommended to clean it each time...

Page 17: ...new fuse Figure 3 reinstall the fuse holder Figure 4 Figure 4 Figure 3 Figure 2 Figure 1 Air Filter Compressor Oxygen and nitrogen separation Adsorption tower Humidi cation bottle Oxygen outlet Oxyge...

Page 18: ...ss the ON OFF button of the operation panel as required 3 Turn on the power switch and restart the device 4 Replace the fuse Firstly cut o the power supply and carry out by professional maintenance pe...

Page 19: ...3 If the phenomenon continues please contact the customer support for consultation see the back cover 7 The inside of the nasal oxygen cannula has a lot of fog or water droplets the method of removing...

Page 20: ...it has been deformed during cleaning and disinfection 2 Replace the disposable kit 3 The cup mouth of the nebulization cup is blocked 3 Clean or replace the nebulization cup 4 Excessive addition of li...

Page 21: ...with the corresponding national laws and regulations It is recommended to consult the environmental protection department or the Omron customer support when dealing with the corresponding waste and r...

Page 22: ...n the manual 2 Portable and mobile radio frequency communication equipment may a ect the normal operation of this product 3 The maximum length of the power cord of this product is 2m 4 In addition to...

Page 23: ...le for use in all establishments including domestic establishments and those directly connected to the public low voltage network that supplies buildings used for domestic purposes Harmonic Emissions...

Page 24: ...the test voltage is applied Floors should be wood concrete or ceramic tile If oors are covered with synthetic material the relative humidity should be at least 30 Mains power quality should be that o...

Page 25: ...he eld strengths should be less than 3V m Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distanc...

Page 26: ...e by maintaining a minimum distance between portable and mobile RF communications equipment trans mitters and this concentrator as recommended below according to the maximum output power of the commun...

Page 27: ......

Page 28: ...he customer support for consultation 5 When performing warranty service if necessary we can request quali ed technicians identi ed by us to provide product circuit diagrams and repairable component da...

Reviews: