6
a. Open battery
compartment
c. Reinstall battery cover
b. Place a new battery inside
Chapter 2 Installing and Setting up the HEA-232
2.1 Inserting / replacing batteries
When should users change the batteries?
If the battery power is too low to perform a test, the
and error
symbol will appear on the display after the device is powered on. In
this case, the system will shut down automatically after 120 seconds.
The battery will have to be replaced before the HEA-232 can be used.
However previously recorded results will not be lost.
There are two conditions which might occur depending on battery
power status
1. The battery power is low:
However, you can still use HEA-232 and results will still be accurate,
but we still strongly suggest you replace the batteries as soon as
possible.
2. The battery symbol flashes on the display after powering on:
The system will shut down automatically after 120 seconds when
only the battery symbol flashes on the display after powering on. It
means the battery power is too low to perform a measurement. You
have to replace the battery in order to continue using the HEA-232;
past results remain recorded.
How to replace HEA-232’s batteries
1. Turn HEA-232 meter over and open the battery cover.
2. Replace one CR-2032 battery with the (+) sign side of battery facing
up.
3. Insert battery cover back to the battery compartment until a
“click”sound is heard.
4. Turn on the device to see whether the HEA-232 is powered on. If
not, please re-examine battery replacing procedures.
5. Please dispose of used batteries according to local regulations.