D. Understanding control test results
When the control solution test result is inside the range on the
printed on the test strip vial, then the HEA-232 and test strips
are working properly.
Caution:The HEA-232 meter measurement unit was set as mg/dL or mmol/L from
manufacturer side. the unit exchange ratio is as the followings: 1 mmol / L = 18.02 mg/dL
If the control solution test result is not within the specified range printed on your test strip vial,
please repeat the control solution test again. If the result continues to fall out of the specified
range, here are some points that need to be checked:
•
Did you follow all of the testing instructions?
•
Are the test strips or control solution expired?
• I
s HEA-232 defective in material or workmanship?
•
Does the code number of the meter matches the code number on the test strip vial?
Repeat the control solution test with a new test strip. If you still have problems, please contact
your local authorized OMRON distributor for more information.
Chapter 5 Memory Function
5.1 Reviewing past results
◆
Press button to turn on the HEA-232, the display will show the most recent test
result then the index and then the most recent test result after a 1 second delay.
Press the button again to display the second recent test result; every time the
button is pressed, it will display a previous result. After the 14th result, it will return
to the first. HEA-232 is able to store up to 14 test results.
Control Solution Range
120 - 150 mg/dL
6.6 - 8.3 mmol/L