19
Sample 2
71.9 mg/dL (4.0mmol/L)
2.7%
Sample 3
134.0 mg/dL (7.4mmol/L)
1.7%
Sample 4
214.4 mg/dL (12mmol/L)
1.5%
Sample 5
396.6 mg/dL (22mmol/L)
1.7%
Between Day Precision
Control solution 1 40.3 mg/dL (2.2mmol/L)
SD=1.5 mg/dL(0.1mmol/L)
Control solution 2 119.8 mg/dL (6.7mmol/L)
1.6%
Control solution 3 328.9 mg/dL (18.3mmol/L)
1.5%
Chapter 9 Specification
◆
Measuring range: 20~600mg/dL (1.1 mmol/L ~ 33.33 mmol/L)
◆
Measuring unit: fixed unit of either mg/dL or mmol/L
◆
Measuring time: Approximately 5 seconds
◆
Sample type: Capillary whole blood samples drawn from the
fingertips
◆
Blood Sample size: Approximately 1μL
◆
Operating temperature: (10
℃
~40
℃
)/(50 ºF ~104ºF)
◆
Relative operating humidity: Below 90% RH
◆
Memory capacity: Up to 14 test results
◆
Dimension: 10 x 3.5 x 2 cm (L×W×H)
◆
Weight: Approximately 35.5g with batteries
◆
Power Supply: One CR-2032 battery
◆
Strip storage condition: at ambient temperature 4
℃
~30
℃
(39
℉
~86
℉
). / below 75% RH
◆
HCT range: 30~55%
warning: The measurement unit either mg/dL or mmol/L is fixed in the meter
according to the country demand. You are not able to change the unit in the
meter.
Caution: Any further service and product information, please contact
OMRON distributors.