P/N 18982-008 Rev A FLOW Service Guide
4
01. Replacing Dongles in the EM2100
Use the following procedure to remove and replace either the EM2100 FLOW License Dongle or the EM2100 Legacy
License Dongle in the EM2100. The EM2100 FLOW License Dongle uses a battery that has a lifespan of
approximately 5 years. Therefore, it is recommended to change the battery every 4 years. Contact the license
administrator to find out how old a dongle is.
CAUTION: ESD DAMAGE RISK.
Protect the appliance from damage caused by electrostatic
discharge (ESD). Wear a grounded wrist or shoe strap whenever removing top cover from the
EM2100.
1. Locations
Parts and tools needed to replace dongles in EM2100 are:
2. Replacement Parts
Part Name
Type
Quantity
Image
1
EM2100 FLOW License Dongle (large black)
14364-200F
1
2
EM2100 Legacy License Dongle (small green)
13131-210F
1
3. Consumables
Name
Note
1
N/A
N/A
4. Necessary Tools
Name
Note
1
Phillips screwdriver
No. 1
Dongles inside of EM2100