Troubleshooting
47
Smoke comms.
error
Smoke Head
reported bad
command mode.
- Reset
- Retry
Smoke head
received bad
command
Contact Omitec
Helpdesk
Communications
fault with the
smoke head.
- Reset
- Retry
Communications
problem with the
smoke head
Check the
handset cable
Fan control error
Fan control
returned error
state
Fans are not
functioning
properly
Contact Omitec
Helpdesk
Tube heater fault
Tube temperature
out of range
Tube temperature
sensor is reporting
a temperature
which is too
low/high
Contact Omitec
Helpdesk
Tube NOT
warming up
The tube
temperature is not
increasing when
the heater is on.
Tube heater may
be damaged.
Contact Omitec
Helpdesk
Contamination
check
Failed check hose
or filter is not
inserted
Contamination
check has failed
Remove hose and
filters
Failed. Please
clean optics
After retrying, the
contaminatin
check has failed
again.
Clean optics. (See
Maintenance
section). If unit
continues to fail,
contact the
Omitec Helpdesk.
Smoke checks
WARNING.
Battery Low.
Connect power
supply to avoid
shutdown.
Battery voltage at
the start of an
MOT test is below
10.5V
Attach external
power supply/
charger.
Print results
Printer Error. Print
results again?
- No
- Yes
Printer has timed
out.
-
Symptom/ Fault
Message
Cause
Action
Summary of Contents for OmiSmoke
Page 2: ......