Technical Data
45
Figure 6-3:
Typical compliance
voltage (50/60 Hz)
The high and low sensitive curves in figure 6-3 correspond to the overload
detection sensitivity settings in the
Test Universe
software. The low sensitive
curves show the maximum available peak compliance voltage, which is
mainly relevant for testing primary and electromechanical relays.
Figure 6-4:
Current derating at high
frequencies for sinusoidal
signals
0.0
20.0
40.0
60.0
80.0
100.0
120.0
140.0
160.0
0
10
20
30
40
50
60
70
80
90
100
Output Current [Arms]
C
o
m
p
li
a
n
ce V
o
lt
ag
e
[V
p
e
ak]
1~ low sensitive 32A (L-L-L-L)
1~ high sensitive 32A (L-L-L-L)
1~ low sensitive 64A (L-L // L-L)
1~ high sensitive 64A (L-L // L-L)
1~ low sensitive 128A (LL-LN)
1~ high sensitive 128A (LL-LN // LL-LN)
6~ low sensitive 32A (L-N)
6~ high sensitive 32A (L-N)
3~ low sensitive 64A (L-N)
3~ high sensitive 64A (L-N)
Output current in A
rms
Com
pliance
voltage
in
V
pe
ak
low sensitive 32 A (L-L-L-L)
high sensitive 32 A (L-L-L-L)
low sensitive 64 A (L-L//L-L)
high sensitive 64 A (L-L//L-L)
low sensitive 128 A (LL-LN)
high sensitive 128 A (LL-LN//LL-LN)
low sensitive 32 A (L-N)
high sensitive 32 A (L-N)
low sensitive 64 A (L-N)
high sensitive 64 A (L-N)
0
4
8
12
16
20
24
28
32
0
100
200
300
400
500
600
700
800
900
1000
Frequency [Hz]
M
a
x.
C
u
rr
en
t
[A
]
Frequency in Hz
Max.
cur
re
nt
in
A
Summary of Contents for CMC 356
Page 1: ...CMC 356 User Manual...
Page 38: ...CMC 356 Reference Manual 38...
Page 100: ...CMC 356 Reference Manual 100...
Page 104: ...CMC 356 Reference Manual 104...
Page 132: ...CMC 356 Reference Manual 132...
Page 136: ...The OMICRON Bootloader Software Copyright Notices 136...