CMC 356 Reference Manual
46
Figure 6-5:
Typical continuous output
current and output power
at 23 °C;
single-phase mode
Figure 6-6:
Typical continuous output
current and output power
at 23 °C;
three- and six-phase mode
The continuous operating range is given by the area below the curves in the
figure 6-5 and 6-6 above.
If you don’t require more than 64 A, we recommend to use the 1 x 64 A
configuration rather than the 128 A one because the 1 x 64 A configuration
provides more continuous output power.
Due to the large number of operating modes, it is not possible to give
universally applicable curves for the discontinuous mode. However, the
examples given below can be used instead to gain feeling for the possible
output durations (t1 is the possible duration of a cold device).
Current in A
C
ontin
uous
o
utput
powe
r
in
W
Current in A
Continuous
output
pow
er
in
W