CMC 356 Reference Manual
82
Typical relative error as function of the excitation
Figure 6-28:
Typical error of the
apparent power S as
function of the excitation,
fs = 28.44 kHz, fin = 50 Hz
Figure 6-29:
Typical error of the active
power P as a function of
the excitation considering
the apparent power,
fs = 28.44 kHz, fin = 50 Hz
Typ. Error apparent pow er S as a function of the excitation
(fs = 28.44kHz, f=50Hz)
0
0,02
0,04
0,06
0,08
0,1
0,12
0,14
0,16
0,18
0,2
0
10
20
30
40
50
60
70
80
90
100
Excitation CH1&CH2 / %
Rel
. Error /
%
phi = 1°
phi = 60°
cos phi=0.01 (89.4°)
Typical error of apparent power S as function of the excitation
(fs = 28.44 kHz, f = 50 Hz)
Rel.
err
or
in
%
Excitation CH1 & CH2 in %
Typ. Error real pow er P as a function of the excitation
(fs = 28.44kHz, f=50Hz)
-0,05
0
0,05
0,1
0,15
0,2
0
10
20
30
40
50
60
70
80
90
100
Excitation CH1&CH2 / %
Rel. Error / %
phi = 1°
phi = 60°
cos phi=0.01 (89.4°)
Typical error of active power P as function of the excitation
(fs = 28.44 kHz, f = 50 Hz)
Rel.
erro
r
in
%
Excitation CH1 & CH2 in %