3.2 Safe Use of the Connecting Cables
3.2.1
Test Lead Adapter for Non-Safety Sockets
The optional CMC Wiring Accessory Package includes flexible test lead
adapters of 5 cm/2 " length with a retractable sleeve (6 x black, 6 x red).
These test leads are to be used as
adapters
, only. They are intended to
make the 4 mm/0.16 " banana plugs of the standard test leads fit into
non-safety sockets (see illustration above).
Never directly insert one of these retractable sleeves into a
CMC 356
output socket at the front of the test set.
This does not comply with the
designated purpose of these leads and is contrary to the safety regulations.
Plug
only the regular test leads
of 2.0 m/6 ft. length into the
CMC 356
output safety sockets.
Retractable sleeve
↔
Safety socket of, for example,
the
CMC 356
test set.
↔
Regular test lead
Non-safety socket
Test lead adapter