10 |
P a g e
4.3.1
2-Point Calibration
A 2-point linearization correction may be applied using either known scaling values or reference weights. 2-Point
Calibration can be performed as described below:
Step 1:
From the
Load Cell Input
interface on SYNC, click
Calibrate
.
Step 2:
Place your low reference and click
Capture
next to
Low Reading
. This will capture the measurement for
the low reference in the low reading box. Enter your actual low reference weight in the
Low Actual
Box.
Step 3:
Place your high reference and click
Capture
next to
High Reading
. This will capture the measurement for
the high reference in the high reading box. Enter your high reference weight in the
High Actual
Box.
Step 4:
Click
Calibrate
to save your calibration.
Note:
Clicking
Calibrate
or
Clear Calibration
will clear the TARE offset.
4.4
Setting Alarms
Alarms are set by clicking the
icon in SYNC on the desired input signal found in the
Inputs Configuration Tab
. Setup
the threshold and alarm type in the
Condition
section and then select which output to turn on in the
Action
section. The
alarm can be set to be latching or non-latching in the
Recovery
section.