33
Configuring The Meter
4
4.12 USING OUTPUT SCALE AND OFFSET (OT.S.O) (Continued)
13.
Press the
T/mV
button. The meter shows the last stored High pH number with
flashing 4th digit.
14.
Press the
/pH
button to change the value of the flashing digit. If you continue to
press the
/pH
button, the flashing digit's value continues to change.
15.
Press the
T/mV
button to scroll to the next digit.
16.
Repeat steps 14 and 15 until the desired value is selected.
17.
Press the
MENU
button to store your selection. The meter shows “
OUT.2
”
(Output 2). (High output)
18.
Press the
T/mV
button. The meter shows selected output.
If you select “
O.2=V
” for voltage, the maximum signal you may select is
10.00 for an 0-10 Vdc signal output. If you select “
O.2=C
” for current,
the maximum signal you may select is 20.00 for a 0-20 or 4-20 mA dc
signal output.
19.
Press the
/pH
button to change the value of the flashing digit. If you continue to
press the
/pH
button, the flashing digit's value continues to change.
20.
Press the
T/mV
button to scroll to the next digit.
21.
Repeat steps 19 and 20 until the desired value is selected.
22.
Press the
MENU
button to store your selection. The meter momentarily shows
“
STRD
”, then “
RST
”. The meter will enter into the “
RUN
” mode (Normal
operation).
CAUTION:
If the meter displays all flashing values on any item, the value
has overflowed. Press the
/pH
button to start new values.