4
Configuring The Meter
26
4.7 USING SETPOINT 2 CONFIGURATION (“S2.CF”)
You may use Setpoint 2 Configuration for the following:
• To set setpoint 2’s active band above or below your chosen value
• To select whether setpoint 2’s operation is latched or unlatched
4.7.1 Setting Setpoint 2's Active Band
1.
Press the
MENU
button until the meter shows “
S2.CF
”.
2.
Press the
T/mV
button. The meter shows one of the following:
•
“S.1=A” (Active above setpoint)
(Default)
• “S.1=B” (Active below setpoint)
3.
Press the
/pH
button to toggle between available choices.
4.
Press the
T/mV
button to select if setpoint 2 is latched or unlatched or press the
MENU
button to store your selection and enter “
S1.DB
” (Setpoint 1 Deadband)
4.7.2 Selecting if Setpoint 2 is Latched or Unlatched
1.
Press the
T/mV
button. The meter shows one of the following:
• “S.2=L” setpoint 2 to be latched
•
“S.2=U” setpoint 2 to be unlatched
(Default)
2.
Press the
/pH
button to toggle between available choices.
3.
Press the
MENU
button to store your choices. The meter momentarily shows
“
STRD
”, followed by “
S1.DB
” (Setpoint 1 Deadband).
4.7.3 Selecting Setpoint 2 (for pH/ORP Controllers only)
1.
Select the
T/mV
button. The meter shows one of the following:
• “S.3=P” Setpoint 2 assigned to pH
• “S.3=O” Setpoint 2 assigned to ORP
2.
Press
/pH/ORP
button to toggle between available choices.
3.
Press the
MENU
button to store your choices. The meter momentarily shows
“
STRD
”, followed by “
S1.DB
” (Setpoint 1 Deadband).