27
Configuring The Meter
4
4.8 SETTING THE SETPOINT 1 DEADBAND (“S1.DB”)
Factory default deadband is 001.0. The deadband’s resolution value depends upon the
decimal point position previously set-up in Section 4.2. The minimum deadband setting
should be no less than 0.1 pH. To change the deadband (hysteresis) of setpoint 1, follow
these steps:
1.
Press the
MENU
button until the meter shows “
S1.DB
”.
2.
Press the
T/mV
button. The meter shows the last stored number (0000 through
9999) with flashing 4th digit.
3.
Press the
/pH
button to change the value of the flashing digit. If you continue to
press the
/pH
button, the flashing digit's value continues to change.
4.
Press the
T/mV
button to scroll to the next digit.
5.
Press the
MENU
button to store your selection. The meter momentarily shows
“
STRD
”, followed by “
S2.DB
” (Setpoint 2 Deadband).
4.9 SETTING THE SETPOINT 2 DEADBAND (“S2.DB”)
Factory default deadband is 001.0. The deadband’s resolution value depends upon the
decimal point position previously set-up in Section 4.2. The minimum deadband setting
should be no less than 0.1 pH. To change the deadband (hysteresis) of setpoint 2, follow
these steps:
1.
Press the
MENU
button until the meter shows “
S2.DB
”.
2.
Press the
T/mV
button. The meter shows the last stored number (0000 through
9999) with flashing 4th digit.
3.
Press the
/pH
button to change the value of the flashing digit. If you continue to
press the
/pH
button, the flashing digit's value continues to change.
4.
Press the
T/mV
button to scroll to the next digit.
5.
Press the
MENU
button to store your selection. The meter momentarily shows
“
STRD
”, followed by “
OT.CF
” (Output Configuration).