25
Configuring The Meter
4
4.6 USING SETPOINT 1 CONFIGURATIONS (“S1.CF”)
You may use Setpoint 1 Configuration for the following:
• To set setpoint 1’s active band above or below your chosen value
• To select whether setpoint 1’s operation is latched or unlatched
4.6.1 Setting Setpoint 1's Active Band
1.
Press the
MENU
button until the meter shows “
S1.CF
”.
2.
Press the
T/mV
button. The meter shows one of the following:
• “S.1=A” (Active above setpoint)
(Default)
• “S.1=B” (Active below setpoint)
3.
Press
/pH
button to toggle between available choices.
4.
Press the
T/mV
button to select if setpoint 1 is latched or unlatched or press the
MENU
button to store your selection and enter “
S2.CF
” (Setpoint 2 configurations).
4.6.2 Selecting if Setpoint 1 is Latched or Unlatched
1.
Select the
T/mV
button. The meter shows one of the following:
• “S.2=L” Setpoint 1 to be latched
•
“S.2=U” Setpoint 1 to be unlatched
(Default)
2.
Press
/pH
button to toggle between available choices.
3.
Press the
MENU
button to store your choices. The meter momentarily shows
“
STRD
”, followed by “
S2.CF
” (Setpoint 2 Configuration).
4.6.3 Selecting Setpoint 1 (for pH/ORP Controllers only)
1.
Select the
T/mV
button. The meter shows one of the following:
• “S.3=P” Setpoint 1 assigned to pH
• “S.3=O” Setpoint 1 assigned to ORP
2.
Press
/pH/ORP
button to toggle between available choices.
3.
Press the
MENU
button to store your choices. The meter momentarily shows
“
STRD
”, followed by “
S2.CF
” (Setpoint 2 Configuration).