Troubleshooting Guide
6
Problem
Solution
If you see an error code, either
“ERR1”, “ERR2”, or “ERR3”, record
the code and call our Customer
Service Department. Provide
Customer Service with the error
code that is displayed in the upper
left corner of the display. The
Customer Service Department
representative may ask you to
return the thermometer to the
factory.
•
The display will go back to the
Real Time Mode (Emissivity
Display Mode).
c. After running the diagnostic
program, press the
key to
unlock the trigger.
1.
The thermometer has to stabilize
before taking temperature
measurements. It takes up to 40
minutes for the thermometer to
stabilize.
1.
The thermometer has to stabilize
before taking temperature
measurements. It takes up to 20
minutes for the thermometer to
stabilize.
☞
The temperature
reading is erratic. The
thermometer has just
been moved from one
extreme temperature to
room temperature
[0°C or 50°C (32°F or
122°F)] or vice versa.
The temperature
reading is erratic. The
thermometer has just
been moved from room
temperature (ambient
temperature) to a
temperature 10°C
colder or warmer.
Summary of Contents for OMEGASCOPE OS520
Page 4: ...ii Unpacking Instructions This page is intentionally blank ...
Page 7: ...v TABLE OF CONTENTS ...
Page 13: ...General Description 1 1 6 Notes ...
Page 49: ...4 3 Sighting Scope 4 Notes ...
Page 57: ...Troubleshooting Guide 6 6 6 Notes ...
Page 63: ...Glossary of Key Strokes 8 8 2 Notes ...
Page 73: ...Appendix Determining an Unknown Emissivity C C 4 Notes ...
Page 77: ...I 4 Index I Notes ...