Product Maintenance

Battery Replacement

Materials: 

OM-CP-BAT110

1.  Unscrew the bottom of the logger and remove the battery.

2.  Place the new battery into the logger. Note the polarity of the battery.

3.  Screw the cover back onto the logger.

Recalibration

The OM-CP-HITEMP140-FP standard calibrations are two points at 30°C and 140°C.

Recalibration is recommended annually for any Omega data logger; a reminder is automatically displayed in 

the software when the device is due.

Notice: Steam Sterilization Applications

The pervasive nature of pressurized steam creates a very difficult environment for electronics.

Please refer to the following preventative maintenance procedure when using this device in steam sterilization 

applications.

Additionally, this device is not ideal for steam sterilization applications above 121 °C/1.1 bar.

PREVENTATIVE MAINTENANCE

After every 3 hours of steam exposure:

1.  Remove the endcap and battery from the device (ref. battery change procedure on Product User Guide)

2.  Place open logger (minus battery) in an oven at 120°C (250°F) for a minimum of 30 minutes

3.  Remove logger from oven and allow to cool to room temp

4.  Re-assemble the logger with the battery (note polarity) and endcap

Quick Start Manual

5

Summary of Contents for OM-CP-HITEMP140-FP

Page 1: ...OM CP HITEMP140 FP High Temperature Data Logger with Flexible RTD Probe e mail info omega com For latest product manuals www omegamanual info Shop online at omega com User sGuide TM...

Page 2: ...ry becomes discharged The OM CP HITEMP140 FP can be configured for delayed start and is capable of reading rates as often as four readings per second up to once every 24 hours Submergibility The OM CP...

Page 3: ...IFC400 is not properly connected 1 In the software click the File button then click Options 2 In the Options window click Communications 3 The Detected Interfaces box will list all of the available c...

Page 4: ...ther parameters appropriate for the data logging application and click Start Quick Start applies the most recent custom start options Batch Start is used for managing multiple loggers at once Real Tim...

Page 5: ...he pervasive nature of pressurized steam creates a very difficult environment for electronics Please refer to the following preventative maintenance procedure when using this device in steam steriliza...

Page 6: ...nds up to 1 reading every 24 hours Required Interface Package OM CP IFC400 or OM CP IFC406 USB docking station required Baud Rate 125 000 baud Typical Battery Life 1 year typical 1 minute reading rate...

Page 7: ...ion or activity or 2 in medical applications or used on humans Should any Product s be used in or with any nuclear installation or activity medical application used on humans or misused in any way OME...

Reviews: