Product Overview
The OM-CP-HITEMP140-FP is a durable, user friendly high temperature data logger featuring a long, flexible
RTD probe with a narrow diameter, making it ideal for use in lyophilization processes.
Commonly used for mapping, validation and monitoring of high temperature surfaces and environments, this
stainless steel data logger is available in two models, the OM-CP-HITEMP140-FP-36 and the
OM-CP-HITEMP140-FP-72, which feature either 36” or 72” flexible probe lengths, respectively. The flexible
probe is coated with PFA insulation and can withstand temperatures up to 260°C with an accuracy of ±0.1°C.
The OM-CP-HITEMP140-FP probe design is narrow and lightweight making it ideal for placement within small
vials, tubing, test tube and other small diameter or delicate applications. Because of the flexible probe, the
risks of breakage (both vial and probe) generally associated with stainless steel probe loggers are diminished
and the location and placement of the probe is easy to manipulate. The device records and stores up to
32,700 time stamped readings and is equipped with non-volatile solid state memory that will continue
recording even if battery becomes discharged.
The OM-CP-HITEMP140-FP can be configured for delayed start and is capable of reading rates as often as four
readings per second, up to once every 24 hours.
Submergibility
The OM-CP-HITEMP140-FP is rated IP68 and is fully submersible. It can be placed in environments up to 230’
(70m) of water.
Bend Radius
The probe should not be bent within 1” where the probe meets the logger or less than 1” from the probe tip.
O-Rings
O-ring maintenance is a key factor when properly caring for the OM-CP-HITEMP140-FP. The O-rings ensure a
tight seal and prevent liquid from entering the inside of the device.
Trigger Settings
The device can be programmed to only record based off user configured trigger settings.
1. In the
Connected devices
panel, select the intended device to change the settings.
2. On the
Device
tab, in the Information group, click
Properties
. Users can also right-click on the device and
select
Properties
in the context menu.
3. Click
Trigger
and configure the
Trigger settings
. Trigger formats are available in Window and Two Point
(bi-level)
mode. Window mode allows for one range of temperature monitoring and two point mode
allows for two ranges.
Note: This product is rated for use up to 140°C. Please heed the battery warning. Additionally, this product is not
ideal for steam sterilization application above 121°C /1.1 bar.
OM-CP-HITEMP140-FP
2