background image

                                                                                                         Temperature/Event Adapter

       4

R

ECORDING 

D

ATA

One of the most important functions of the Service Logger is to record
information over a period of time. This function is like a chart recorder except
that the information is stored in memory rather than being printed on paper.
The user can choose to print the information, transfer the information to a
computer for advanced analysis or long term storage, or review the information
on the Service Logger’s display.

In the case of the OM-60-MOD-TE sensor, the temperature and presence or
absence of voltage will be recorded. Different sensors will, of course, record
different information depending on the function of that sensor. The Service
Logger has a memory capacity to record up to 4000 samples of data. It is the
option of the user to record data to the maximum capacity of the Service Logger
or to terminate a recording when sufficient information has been gathered.

IMPORTANT !
Each new recording will erase any previously recorded information. It is the
responsibility of the user to either print or store a previous recording prior
to starting a new recording.

The first step in making a recording is to select the proper recording time and
sample rate. If a sampling rate of 15 seconds is being used, then every 15
seconds the temperature and presence or absence of voltage will be measured
and stored into digital memory. No record of information is made between each
sample even though the display will continuously show the environment the
OM-60-MOD-TE sensors are exposed to.

If  the Service Logger is set to record on event change, then only when the state
of the voltage detector goes from ON to OFF or from OFF to ON will a data
recording be stored in memory.

For a recording to accurately reflect a series of events, it is up to the user to
select a recording interval that best suits the particular application. In some
situations maximum recording time may be sacrificed so that more frequent
sampling may be accomplished. Long term monitoring will require less frequent
sampling in favor of longer recording times. The record on event change will
usually be used when the data of interest occurs only under a specific
condition.

The next step is to select how the Service Logger will record the information
from the OM-60-MOD-TE sensor. There are 3 possible ways to record
information.

Summary of Contents for OM-60-MOD-TE

Page 1: ...D TE cables by the user is not recommended and will void the sensor warranty The voltage detector is connected by two coiled clip leads available at the interface box The voltage detector will operate...

Page 2: ...s not connected or has become defective the display will indicate that the sensor is not responding Temperature sensor is not responding CHANGING SETTINGS Every function of the Service Logger can be s...

Page 3: ...r and event detector but will only record data on a change in the state of the event detector When the M ENU EN TER button is pressed 3 more times to cycle through the remaining functions the recordin...

Page 4: ...ding prior to starting a new recording The first step in making a recording is to select the proper recording time and sample rate If a sampling rate of 15 seconds is being used then every 15 seconds...

Page 5: ...minate recording if the Service Logger was set to record previously 3 Record information to memory and to a printer This is a combination of options 1 and 2 Information is stored in memory and at the...

Page 6: ...the upper right hand corner of the display The R indicates recording is in progress While you are recording it is possible to check how much memory has been used Press and hold the STA TU S button unt...

Page 7: ...s is that data samples that are less than the selected temp erature change or if the state of the event detector stays the same no samples will be printed until the conditions that are programmed in t...

Page 8: ...he temperature continued to decrease with the fan OFF at 5 32 a m on the Tuesday of the same week End of Data This indicates the end of the recording and shows that the temperature remained within 2 o...

Page 9: ...to operate an error message will be displayed to alert the user to a printer problem There are several possible problems that may occur if the printer is not ready The Service Logger will alert the u...

Page 10: ...have to manually scan a potential 4000 data samples to find the information of interest Press and hold M ENU EN TER until this display appears The actual recording rate is unimportant Press M ENU EN...

Page 11: ...g of the data recording or R EC O R D to go to the last previous data sample Press M ENU EN TER at any time to return to the Operating Mode TRANSFERRING DATA TO A COMPUTER After a recording has been m...

Page 12: ...puter screen A README file is also included on the floppy diskette which has detailed instructions on the use of the transfer program and reflects the latest information that is not included in this m...

Reviews: