background image

5

Figure 5

Power Input: 

(See Figure5)

The LV800 is designed to accept either 115VAC or 230VAC
(factory set).  Verify the intended voltage supply is compatible
with the voltage configuration indicated on the electronics and
the external nameplate.  Connect power as shown in Figure 5.
Select wire size that can deliver suitable voltage and current for
the application.

Output Relay Contacts:  

(See Figure 5)

The LV800 is equipped with one or two sets of isolated contacts
which indicates whether or not material is being detected with-
in the vessel.  This output is also influenced by the selection of
the "fail-safe" switch as described in the "Calibration" section of
this manual.  The designations on the circuit board relate to the
contact status when the material is "not" sensed and the fail-
safe switch is in the "low" mode.  These contacts can be con-
nected to any type of control device, provided that ratings are
observed (See Specifications).  Select wire size that can deliv-
er suitable voltage and current for the application.

C A L I B R AT I O N

Sensitivity:

(See Figure 6)

Sensitivity is set in the field via a "Sense Adjust" 25-turn poten-
tiometer.  

The “Sense Adjust” tunes the sensitivity of the LV800 to the
material being sensed.  Rotating the adjustment in a clockwise
direction will increase the sensitivity making the material easier
to “detect”.  There are two different procedures which can be
used in the calibration process.  Both utilize the simple “2-Step
Calibration” procedure.  One procedure sets the sensitivity with
the probe being exposed to the material, while the second pro-
cedure exposes the probe to free air only.  Since the capaci-
tance being measured is directly related to the material being
sensed, the best results will be achieved by setting the calibra-
tion while the probe is exposed to the application material. 

2-Step Calibration- probe exposed to free air only
1)

Rotate “Sense Adjust” clockwise just until the “Sense” 
(yellow) LED turns on.

2)

Reposition “Sense Adjust” a number of turns counter-
clockwise with respect to the three described sensitivities 
below (see Figure 7).  It is desirable to reposition “Sense 
Adjust” in accordance to “2-Step Calibration - probe 
exposed to material” procedure once material can cover the
probe. Materials with high dielectric constants or conductiv-
ity, require the sensitivity adjustment further counter-
clockwise than those applications with low dielectric 
constants.

2-Step Calibration- probe exposed to material
1)

While the probe is not covered with material, rotate “Sense
Adjust” clockwise just until “Sense” (yellow) LED turns on.

2)

Permit the application material to cover the probe.  

While 

counting the number of revolutions

, rotate “Sense  

Adjust” counter-clockwise until the “Sense” (yellow) LED 
turns off. Reposition “Sense Adjust” at the half way point 
between the LED on and LED off settings.  (e.g. After insert-
ing the probe into the material, if it takes 2 counter-clockwise
turns of “Sense Adjust” for the “Sense” (yellow) LED to turn
off, then “Sense Adjust” should be repositioned 1 clockwise
turn.)

Figure 6

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for LV800 SERIES

Page 1: ...omega com e mail info omega com For latest product manuals omegamanual info LV800 SERIES Capacitance Point Level Sensors Shop online at User sGuide...

Page 2: ...8280 Guyancourt France TEL 33 0 1 61 37 2900 FAX 33 0 1 30 57 5427 Toll Free in France 0800 466 342 e mail sales omega fr Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany TEL 49 0 7056 9...

Page 3: ...of this manual pertaining to SAFETY Doing so will ensure proper oper ation of the equipment and the safety of all personnel Before discarding shipping container please inspect it thoroughly and verify...

Page 4: ...for at least 8 to 12 inches 203 305 mm of probe coverage When sensing highly conductive materials or those with a high dielectric constant the LV800 can perform as a tip sensitive sensor 2 Vessel Con...

Page 5: ...tion Doing will likely cause improper operation of the sensor Permanently Connected Equipment Disconnecting devices shall by included in the system installa tion In installations were multiple circuit...

Page 6: ...th utilize the simple 2 Step Calibration procedure One procedure sets the sensitivity with the probe being exposed to the material while the second pro cedure exposes the probe to free air only Since...

Page 7: ...tions are reversed when no material is sensed and high fail safe is selected Relay is energized Figure 7 LV800 Indicators See Figure 9 1 Yellow LED Its status describes the sensing condition of the LV...

Page 8: ...nd when no material is sensed If Fail safe High mode is used the designations are reversed Swap wire terminations of N O and N C if necessary Changing the fail safe selection is not recommended Fuse R...

Page 9: ...8 MECHANICALS Standard Probe Stub Probe DIMENSIONS ARE SHOWN IN INCHES WITH MILLIMETER EQUIVALENT IN BRACKETS...

Page 10: ...essential that all instructions in this manual be followed to ensure proper operation of the equipment and safety of operating person nel Use of equipment not specified herein may impair protection p...

Page 11: ...erwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold by...

Page 12: ...bine Paddlewheel Systems M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U Industri...

Reviews: