The figure below shows Voltage (bridge with amplified output) input with internal
excitation.
Figure 2.8 4-Wire Voltage Input
(Bridge with Amplified Output) with Internal Excitation
Where: +S: signal plus, -S: signal return, +Ext: excitation plus, -Ext: excitation return
+E: plus excitation sense, -E: minus excitation sense.
This meter is capable of supplying 5 or 10 Vdc sensor excitation. The excitation
output connection and location of S2 pin selection jumper are shown below.
If your meter has an excitation option, then communication is not available.
Figure 2.9
a) Excitation Output
b) Top View Location of S2
Install jumpers according to the table below.
Table 2.3 Jumper Connections
Factory default is 10 V.
10
Excitation Output
S2
A
B
10 V
Close
Open
5 V
Open
Close