8 of 36
1.
To zero the manometer using
Referenced to Absolute Zero
,
start with the unit turned
OFF
and use the following
keystroke sequence:
Keystroke
Display
1. Press ON/OF button.
The display briefly shows the
header name and full scale range
of the unit in the last engineering
units selected. The manometer
then goes into the
Measure
Mode
where the applied pressure
and engineering unit of measure
are displayed.
2. Connect the HHP360 to
a vacuum source capable
of a vacuum of 100
microns absolute pressure
or less.
3. Pull a full vacuum.
Display should read close to
zero. (See note on next page)
4. Press MIN/MAX and
TARE keys at the same
time. (See figure 1 below.)
Top line of display reads
“ZEROING SOURCE:”
Bottom line of display reads
“REF TO ABS ZERO”
5. Press the PRGM key.
Top line of display reads “ZERO
IN PROGRESS” while bottom
line counts down from 9. Zeroing
is complete when unit returns to
Measure Mode.
Figure 1
Summary of Contents for HHP360 Series