18 of 36
6.
To turn on the correction,
set ENABLED on the 2
nd
line & press the PRGM
key.
-OR-
To turn off the correction,
set DISABLED on the
2nd line and skip ahead to
step 12.
Top line reads “VALUE=:
00000000”. Bottom line reads
“CHANGE?:
NO”.
7.
To change the value press
the ▲ key to toggle the
display to “YES”.
Top line reads “VALUE=:
00000000”. Bottom line reads
“CHANGE?:
YES”.
8.
Press the PRGM key.
Top line reads “SEA LEVEL
METERS”. Bottom line reads
“00000000”.
9.
Press the ▲ or ▼ arrow
keys to set the correct
value in the first digit.
Top line reads “SEA LEVEL
METERS”. Bottom line reads
“70000000”.
10.
When the value is correct
press the PRGM key.
Cursor moves over to the next
digit.
11.
Repeat steps 9 and 10
above until the correct
elevation is entered.
Top line reads “SEA LEVEL
METERS”. Bottom line reads
“700.0000”.
12.
Press the PRGM key to
continue moving the
cursor to the right. When
the last digit is entered the
unit will leave the sea
level select mode and
return to the Program
Mode.
Top line reads “PROGRAM
MODE”. Bottom line reads
“UNITS SELECT”.
13.
Press the ◄ key. The
display will return to the
Measure Mode. When
the Correction to Sea
Level is ENABLED, the
display will have the
letter “S” at the
beginning of the 2nd
line.
Display reads;
“S 803.1 TORR”