Operating Instructions
3-6
The first line of the display indicates the current Channel 1 temperature, regardless of
which view mode or channel’s statistic is currently displayed.
M
ODEL
C
HANNEL
S
TATISTIC
V
IEW
S
EQUENCE
HH931T
T1
MIN
MAX
AVG
RNG
STDEV
HH932T
T1
MIN
MAX
AVG
RNG
STDEV
T2
MIN
MAX
AVG
RNG
STDEV
Figure 8: Statistics Sequence
If the instrument records invalid measurement data during the statistics session such as an
over-range, under-range, or open input value,
———— will be displayed for each affected statistic result.
To return to the active measurement mode, press
repeatedly to step through the
remaining view modes, or cycle power.
Auto-Power Off
Key designators followed by (1.5s), e.g.
(1.5s)
,
indicate that the key should be
pressed and held for 1.5 seconds, then released to access the desired function.
To conserve battery life, the instrument automatically turns off if no key is pressed for 20
minutes. To disable this feature, press
. The remaining battery life indicator will flash
once, indicating auto-power off is disabled.
Auto-power off will remain disabled until instrument power is cycled. At next power on,
auto-power off returns to the default enabled condition.
While data collection is active, the auto-power off feature is automatically disabled to ensure
uninterrupted data collection.
Backlight and Backlight Timeout
The instrument includes an LED backlight feature to ensure measurement data can be easily
read in low-light conditions. To activate the backlight, press
.
Once the backlight is activated, it will automatically turn off after 30 seconds if no key is
pressed to preserve battery life. To disable the backlight timeout feature, press
(1.5s)
. The
backlight will flash to indicate the timeout feature has been disabled. To re-enable the
backlight timeout feature, turn the backlight off then on by pressing
twice.
The backlight feature cannot be activated or deactivated from the OMEGA Link mobile app.
Summary of Contents for HH931T
Page 39: ...Appendices B i B EXPANDED INSTRUMENT UNCERTAINTIES Thermocouple Type B ...
Page 40: ...Appendices B ii Thermocouple Type E ...
Page 41: ...Appendices B iii Thermocouple Type J ...
Page 42: ...Appendices B iv Thermocouple Type K ...
Page 43: ...Appendices B v Thermocouple Type N ...
Page 44: ...Appendices B vi Thermocouple Type R ...
Page 45: ...Appendices B vii Thermocouple Type S ...